Rule 37.14.1307 - OPERATING REQUIREMENTS
(1) Normal operating procedures shall be written and made available to all analytical x-ray equipment workers. No person shall be permitted to operate analytical x-ray equipment in any manner other than that specified in the procedures unless such person has obtained written approval of the radiation safety officer.
(2) No person shall bypass a safety device unless such person has obtained the approval of the radiation safety officer. Such approval, if given, shall be for a specified period of time. When a safety device has been bypassed, a readily discernible sign bearing the words "SAFETY DEVICE NOT WORKING", or words having a similar intent, shall be placed on the radiation source housing.