Suction aspirator system: Shall be electrically powered. Shall provide a free airflow of at least 30 lpm at the distal end of the connected patient hose. It shall achieve a vacuum of at least 300 mmHG (11.8 inches) within 4 seconds after the suction tube is clamped closed.