The requirements of this part are in addition to other applicable requirements of this chapter.
All radiographic operations must be conducted in a permanent radiographic installation, unless specifically authorized by the commissioner.
The control panel of each radiation-producing machine must be equipped with a locking device that will prevent the unauthorized use of an x-ray system or the accidental production of radiation. The radiation-producing machine must be kept locked and the key removed at all times except when under the direct visual surveillance of a radiographer.
Radiation-producing machines must be secured while in storage to prevent tampering or removal by unauthorized individuals.
An entrance that is used for personnel access to the high radiation area in a permanent industrial radiographic installation must meet the requirements of part 4732.0620, subpart 1.
Registrants must provide individual monitoring devices according to part 4732.0440. At permanent radiographic installations where alarming or warning devices are in routine use, the use of a direct reading dosimeter is not required.
Registrants must maintain records of alarm system and entrance control device tests required under this part and retain each record according to part 4732.0330.
Minn. R. agency 144, ch. 4732, RADIATION THERAPEUTIC REQUIREMENTS, pt. 4732.1050