Part 4731.4513 - BRACHYTHERAPY SOURCE CALIBRATION RECORDS A licensee must maintain a record of the calibrations of brachytherapy sources required under part 4731.4455 for three years after the last use of the source. The records must include:
A. the date of the calibration;B. the manufacturer's name, the model number and serial number for the source, and instruments used to calibrate the source;C. the source output or activity;D. the source positioning accuracy within the applicators; andE. the signature of the authorized medical physicist.Minn. R. agency 144, ch. 4731, MEDICAL USE OF RADIOACTIVE MATERIAL, pt. 4731.4513