A licensee providing mobile remote afterloader service must:
In addition to the periodic spot checks required under part 4731.4473, a licensee authorized to use mobile afterloaders for medical use must perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks must be made to verify the operation of:
In addition to the requirements for checks under subpart 2, a licensee must ensure proper overall operation of the remote afterloader unit by conducting a simulated cycle of treatment before use at each address of use.
If the results of the checks required under subpart 2 indicate the malfunction of any system, a licensee must lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
A licensee must retain a record of each check required under subpart 2 according to part 4731.4522.
Minn. R. agency 144, ch. 4731, MEDICAL USE OF RADIOACTIVE MATERIAL, pt. 4731.4475
Statutory Authority: MS s 144.1202; 144.1203