A licensee authorized to use remote afterloader units for medical use must perform spot checks of each remote afterloader facility and on each unit:
A licensee must perform the measurements required under subpart 1 according to written procedures established by the authorized medical physicist. The authorized medical physicist need not actually perform the spot check measurement.
A licensee must have the authorized medical physicist review the results of each spot check within 15 days. The authorized medical physicist must notify the licensee as soon as possible in writing of the results of the spot check.
To satisfy subpart 1, spot checks must, at a minimum, ensure proper operation of:
If the results of the checks required under subpart 4 indicate the malfunction of any system, a licensee must lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
A licensee must retain a record of each check required under subpart 4 and a copy of the procedures required under subpart 2 according to part 4731.4520.