Minn. R. agency 144, ch. 4731, MEDICAL USE OF RADIOACTIVE MATERIAL, pt. 4731.4455

Current through Register Vol. 49, No. 29, 2025
Part 4731.4455 - BRACHYTHERAPY; CALIBRATION MEASUREMENTS
A. Before the first medical use of a brachytherapy source, a licensee must have:
(1) determined the source output or activity using a dosimetry system that meets the requirements of part 4731.4468, subpart 1;
(2) determined source positioning accuracy within applicators; and
(3) used published protocols currently accepted by nationally recognized bodies to meet the requirements of this item.
B. A licensee may use measurements provided by the source manufacturer or by a calibration laboratory accredited by the American Association of Physicists in Medicine that are made according to item A.
C. A licensee must mathematically correct the outputs or activities determined under item A for physical decay at intervals consistent with one percent physical decay.
D. A licensee must retain a record of each calibration according to part 4731.4513.

Minn. R. agency 144, ch. 4731, MEDICAL USE OF RADIOACTIVE MATERIAL, pt. 4731.4455

29 SR 755

Statutory Authority: MS s 144.1202; 144.1203