Upon first entering the radiation room of a panoramic irradiator after an irradiation, the irradiator operator must use a survey meter to determine that the source has returned to its fully shielded position. The operator must check the functioning of the survey meter with a radiation check source before entry.
Before exiting from and locking the door to the radiation room of a panoramic irradiator prior to a planned irradiation, the irradiator operator must:
During a power failure, the area around the pool of an underwater irradiator may not be entered without using an operable and calibrated radiation survey meter, unless the over-the-pool monitor required under part 4731.6090, subpart 2, is operating with backup power.
Minn. R. agency 144, ch. 4731, IRRADIATORS, pt. 4731.6240
Statutory Authority: MS s 144.1202; 144.1203