Each dry-source-storage sealed source must be tested for leakage at intervals not to exceed six months using a leak test kit or method approved by the commissioner, the NRC, or an agreement state. In the absence of a certificate from a transferor that a test has been made within the six months before the transfer, the sealed source may not be used until tested. The test must be capable of detecting the presence of 0.005 microcurie (200 becquerels) of radioactive material and must be performed by a person approved by the commissioner, the NRC, or an agreement state to perform the test.
Minn. R. agency 144, ch. 4731, IRRADIATORS, pt. 4731.6200
Statutory Authority: MS s 144.1202; 144.1203