Mich. Admin. Code R. 333.5365

Current through Vol. 24-24, January 15, 2025
Section R. 333.5365 - Enclosures

Rule 365.

(1) The degree of protection required for an enclosure shall be determined by the workload, use and occupancy factors, and the tube potential, tube current, mechanical movement, and distance, and shall be subject to design approval by the department.
(2) In a radiographic room, wall and floor areas exposed to the useful beam plus an area extending at least 30 centimeters (1 foot) beyond shall be provided with a primary protective barrier where necessary as determined by workload, use and occupancy factors, and distance.
(3) Secondary protective barriers shall be provided in the radiographic room ceiling and in those walls not requiring primary barriers. Common building materials often fulfill this requirement.
(4) A fixed barrier of 1.6 millimeters (1/16 inch) lead equivalence, such as a shielded wall partition or immobilized portable shield, is recommended for operator protection. When this protection is provided, the operator shall be able to see and communicate with the patient from a shielded position.

Mich. Admin. Code R. 333.5365

2016 MR 10, Eff. 5/25/2016