Mich. Admin. Code R. 333.5694

Current through Vol. 24-24, January 15, 2025
Section R. 333.5694 - Mammography equipment evaluations

Rule 694.

(1) Additional evaluations of stereotactic breast biopsy units or image processors shall be conducted when a new unit or processor is installed, a unit or processor is disassembled and reassembled at the same or a new location, or major components of a stereotactic breast biopsy unit or processor equipment are changed or repaired. These evaluations shall be used to determine whether the new or changed equipment meets the requirements of R 333.5676 to R 333.5678 and R 333.5687 to R 333.5691, as applicable. Problems revealed by the evaluation shall be corrected before the new or changed equipment is put into service for procedures or film processing.
(2) The equipment evaluations specified in subrule (1) of this rule shall be performed by a medical physicist or by an individual under the direct supervision of a medical physicist.

Mich. Admin. Code R. 333.5694

2016 MR 10, Eff. 5/25/2016