Mich. Admin. Code R. 333.5689

Current through Vol. 24-24, January 15, 2025
Section R. 333.5689 - Phantom image scores

Rule 689. A stereotactic breast biopsy phantom image score for the tests required in R 333.5687(b) and R 333.5688(h) shall be not less than the values specified in table 689:

TABLE 689

Image System

Standard Mammography Phantom

Mini Stereotactic Phantom

Fibers

Speck Groups

Masses

Fibers

Speck Groups

Masses

Screen-film

4.0

3.0

3.0

2.0

2.0

2.0

Digital

5.0

4.0

3.5

3.0

3.0

2.5

Mich. Admin. Code R. 333.5689

2016 MR 10, Eff. 5/25/2016