Current through Register Vol. 52, No. 1, January 10, 2025
Section 10.10.03.06 - Letter of Exception-Limited Testing for Rare Diseases Standards and RequirementsA. Application Requirements and Testing Approval. A person required to obtain a license to operate a laboratory that chooses to operate under a letter of exception-limited testing for rare diseases shall: (1) Submit: (a) A completed: (i) Application as prescribed in Regulation .04 of this chapter; and(ii) Report or questionnaire as required by the Secretary to determine regulatory compliance;(b) A copy of the laboratory's:(i) Current CLIA certificate issued by CMS; and(ii) Most recent CLIA survey report;(c) Documentation of the rare disease testing offered or performed, including but not limited to: (i) The test procedure, including instructions for quality control and quality assurance;(ii) A description of the disease or condition;(iii) An explanation of the clinical validity of the test;(iv) The performance characteristics of the test;(v) The expected turnaround time for patient test results; and(vi) Additional information necessary for determining eligibility and compliance with this subtitle; and (2) Agree to: (a) Make records and reports available to the Department as the Secretary may require;(b) Notify the Department when a change to a licensee's rare disease testing procedure occurs for the purpose of the Department's determining whether the change is in compliance with this subtitle;(c) Perform not more than 50 rare disease tests each year on specimens from Maryland patients; and(d) Submit to the Department statistics that detail the number and types of rare disease tests performed each year on specimens from Maryland patients.B. Standards. A person that wishes to obtain and maintain a letter of exception-limited testing for rare diseases shall: (1) Comply with good laboratory practices that include:(a) Following the manufacturer's instructions for instrument or test system operation and test performance;(b) Having and following a written procedure manual describing the processes for performing tests and reporting patient test results;(c) Performing applicable instrument calibration procedures at least once every 6 months;(d) Performing and documenting quality control procedures using applicable levels of control materials each day of testing; and(e) Performing and documenting that remedial action has been taken when problems or errors are identified; and(2) Meet the requirements and standards in this subtitle applicable to a permitted laboratory including: (a) Proficiency testing under COMAR 10.10.05;(b) Quality assurance under COMAR 10.10.06;(c) Personnel under COMAR 10.10.07;(d) Sanctions under COMAR 10.10.08; and(e) CLIA regulations incorporated by reference at COMAR 10.10.01.04B(3)-(5).C. If the submitted application and explanatory documentation do not meet the requirements of this regulation or if the information submitted for the plan is insufficient to make a determination of compliance with this subtitle, a person may: (1) Submit amendments to the application and explanatory documentation; or(2) Apply for a permit in accordance with this subtitle.D. The OHCQ may conduct on-site surveys of a laboratory operating under a letter of exception-limited testing for rare diseases to:(1) Investigate a complaint; or(2) Determine compliance with this subtitle.E. If a licensee wishes to perform more than 50 rare disease tests per year, the licensee shall submit an application for and obtain a permit as required under this subtitle.Md. Code Regs. 10.10.03.06
Regulation .06 adopted effective June 1, 2009 (36:11 Md. R. 787); amended effective 43:17 Md. R. 953, eff. 8/29/2016