10-144-220 Me. Code R. § G-L-2632

Current through 2024-51, December 18, 2024
Section 144-220-G-L-2632 - Records of teletherapy, remote afterloader, and gamma stereotactic radiosurgery full calibrations
A. A licensee shall maintain a record of the teletherapy unit, remote afterloader unit, and gamma stereotactic radiosurgery unit full calibrations required by G.632, G.633, and G.635 for three years.
B. The record must include:
(1) The date of the calibration;
(2) The manufacturer's name, model number, and serial number of the teletherapy, remote afterloader, and gamma stereotactic radiosurgery unit(s), the source(s), and the instruments used to calibrate the unit(s);
(3) The results and an assessment of the full calibrations;
(4) The results of the autoradiograph required for low dose-rate remote afterloader units; and
(5) The signature of the authorized medical physicist who performed the full calibration.

10-144 C.M.R. ch. 220, § G-L-2632