Current through 2024-51, December 18, 2024
Section 144-220-G-H-655 - Full-Inspection Service for teletherapy and gamma stereotactic radiosurgery unitsA. A licensee shall have each teletherapy unit and gamma stereotactic radiosurgery unit fully inspected and serviced during source replacement to assure proper functioning of the source exposure mechanism and other safety components. The interval between each full inspection shall not exceed five years for each teletherapy unit and shall not exceed seven years for each gamma stereotactic radiosurgery unit.B. This inspection and servicing may only be performed by persons specifically licensed to do so by the Agency, the Nuclear Regulatory Commission, or an Agreement State.C. A licensee shall keep a record of the inspection and servicing in accordance with G.2655.10-144 C.M.R. ch. 220, § G-H-655