Current through 2025-03, January 15, 2025
Section 144-220-G-H-632 - Full calibration measurements on teletherapy unitsA. A licensee authorized to use a teletherapy unit for medical use shall perform full calibration measurements on each teletherapy unit: (1) Before the first medical use of the unit; and(2) Before medical use under the following conditions: (a) Whenever spot-check measurements indicate that the output differs by more than 5 percent from the output obtained at the last full calibration corrected mathematically for radioactive decay;(b) Following replacement of the source or following reinstallation of the teletherapy unit in a new location;(c) Following any repair of the teletherapy unit that includes removal of the source or major repair of the components associated with the source exposure assembly; and(3) At intervals not exceeding one year.B. To satisfy the requirement of G.632.A, full calibration measurements must include determination of:(1) The output within +/-3 percent for the range of field sizes and for the distance or range of distances used for medical use;(2) The coincidence of the radiation field and the field indicated by the light beam localizing device;(3) The uniformity of the radiation field and its dependence on the orientation of the useful beam;(4) Timer accuracy and linearity over the range of use;(6) The accuracy of all distance measuring and localization devices in medical use.C. A licensee shall use the dosimetry system described in G.630.A to measure the output for one set of exposure conditions. The remaining radiation measurements required in G.630.B.(1) may be made using a dosimetry system that indicates relative dose rates.D. A licensee shall make full calibration measurements required by G.632.A in accordance with published protocols accepted by nationally recognized bodies.E. A licensee shall mathematically correct the outputs determined in G.632.B.(1) for physical decay for intervals not exceeding one month for cobalt-60, six months for cesium-137, or at intervals consistent with 1 percent decay for all other nuclides..F. Full calibration measurements required by G.632.A and physical decay corrections required by G.632.E must be performed by the authorized medical physicist.G. A licensee shall retain a record of each calibration in accordance with G.2632.10-144 C.M.R. ch. 220, § G-H-632