In addition to other requirements of this part, all diagnostic x-ray systems shall meet the following requirements:
TABLE 1
Minimum HVL (mm of AL) | |||
Design operating range (Kilovolts peak) | Measured potential (kVp) | Specified dental systems* | All other diagnostic X-ray systems |
Below 50 --------- | 30 | 1.5 | 0.3 |
40 | 1.5 | 0.4 | |
49 | 1.5 | 0.5 | |
50 to 70 --------- | 50 | 1.5 | 1.2 |
60 | 1.5 | 1.3 | |
70 | 1.5 | 1.5 | |
Above 70 --------- | 71 | 2.1 | 2.1 |
80 | 2.3 | 2.3 | |
90 | 2.5 | 2.5 | |
100 | 2.7 | 2.7 | |
110 | 3.0 | 3.0 | |
120 | 3.2 | 3.2 | |
130 | 3.5 | 3.5 | |
140 | 3.8 | 3.8 | |
150 | 4.1 | 4.1 | |
* for any dental x-ray system designed for use with intraoral image receptors and manufactured after December 1, 1980; |
TABLE 2
Filtration Required vs. Operating Voltage | |
Total Filtration (inherent plus added) | |
Operating Voltage (kVp) | (millimeters aluminum equivalent) |
Below 50 ----------------------------------- | 0.5 |
50 - 70 -------------------------------------- | 1.5 |
Above 70 ----------------------------------- | 2.5 |
10-144 C.M.R. ch. 220, § F-5