Authority: IC 16-41-35-26; IC 16-41-35-29
Affected: IC 16-41-35
Sec. 127.
The statement must further indicate that the original images or films are being provided to the mammography facility for inclusion in the individual's medical record.
Compliance shall be based on determination of the coefficient of variation and the average of at least four (4) consecutive measurements for the kVp at which the x-ray equipment is normally used. Compliance may be based on single measurements for other kVps over the range of use. However, if any single measurement is out of compliance, an average and coefficient of variation shall be determined at that kVp for comparison to subdivisions (1) and (2).
[X1 - X2] <=0.10 [ X1 + X2]
Where: X1 and X2 = The ratios of average exposure to mAs obtained at each of two (2) consecutive tube current settings
TABLE V
Parameter | Frequency | Individual Responsible |
Darkroom cleanliness | daily | mammographer |
Processor performance | daily | mammographer |
Screen cleanliness | weekly | mammographer |
View boxes and viewing conditions | weekly | mammographer |
Image quality (phantom images) | monthly | mammographer |
Repeat analysis | quarterly | mammographer |
Analysis of fixer retention in film | quarterly | mammographer |
Darkroom fog | semiannually | mammographer |
Screen-film contact | semiannually | mammographer |
Compression | semiannually | mammographer |
AEC density control function | annually | diagnostic imaging physicist |
Star pattern focal spot size test | annually | diagnostic imaging physicist |
Uniformity of screen speed | annually | diagnostic imaging physicist |
Assembly physical evaluation | annually | diagnostic imaging physicist |
If the results fall outside the acceptable range, the test must be repeated. If the results continue to be unacceptable, the cause of the problem must be identified and corrected before further examinations are conducted.
410 IAC 5-6.1-127