Authority: IC 16-41-35-26; IC 16-41-35-29
Affected: IC 16-41-35
Sec. 119.
Mammography facilities shall be evaluated at least once each twelve (12) months by a diagnostic imaging physicist approved by the department. All other diagnostic x-ray systems shall be evaluated at least once each twelve (12) months by a diagnostic imaging physicist or x-ray machine inspector approved by the department. Those x-ray facilities which have been evaluated within sixty (60) days after the end of the interval established in this section will be considered to be in compliance with this section as long as the evaluation occurs in the same calendar year as the date on which reevaluation is required. All diagnostic x-ray systems shall comply with this section.
TABLE I
Design Operating Range | Measured Voltage | Half-Value Layer Aluminum Equivalent |
Below 50 kVp | 30 kVp | 0.3 mm |
40 kVp | 0.4 mm | |
49 kVp | 0.5 mm | |
50 to 70 kVp | 50 kVp | 1.2 mm |
60 kVp | 1.3 mm | |
70 kVp | 1.5 mm | |
71 kVp | 2.1 mm | |
80 kVp | 2.3 mm | |
90 kVp | 2.5 mm | |
100 kVp | 2.7 mm | |
110 kVp | 3.0 mm | |
120 kVp | 3.2 mm | |
130 kVp | 3.5 mm | |
140 kVp | 3.8 mm | |
150 kVp | 4.1 mm |
For a kVp not listed in Table I, linear interpolation shall be utilized to determine the minimum acceptable half-value layer. For capacitor energy storage x-ray systems, compliance shall be determined with the maximum charge per exposure assumed to be the kVp. The required minimum aluminum equivalent filtration shall include the filtration contributed by all materials which are always present between the source and the patient. The requirements of this subsection will be considered to have been met if it can be demonstrated that the aluminum equivalent of the total filtration in the primary beam is not less than that shown as follows:
TABLE II
Total Filtration | |
Operating Voltage | Aluminum Equivalent |
Below 50 kVp | 0.5 mm |
50 to 70 kVp | 1.5 mm |
Above 70 kVp | 2.5 mm |
In addition, there must be compliance with the following:
410 IAC 5-6.1-119