410 Ind. Admin. Code 5-5-7

Current through January 8, 2025
Section 410 IAC 5-5-7 - Survey instruments for gamma and temporary job site radiography

Authority: IC 16-41-35-26; IC 16-41-35-29

Affected: IC 16-41-35

Sec. 7.

(a) The licensee or registrant shall maintain sufficient calibrated and operable radiation survey instruments to make physical radiation surveys as required by 410 IAC 5-5-7 and 410 IAC 5-4-9 of 410 IAC 5. Instrumentation required by this section shall have such a range such that 2 milliroentgens per hour through 1 roentgen per hour can be measured.
(b) Each radiation survey instrument shall be calibrated:
(1) at energies appropriate for use and at intervals not to exceed 3 months and after each instrument servicing;
(2) such that accuracy within plus or minus 20 percent traceable to a national standard can be demonstrated; and
(3) at two or more widely separated points, other than zero, on each scale.
(c) Records of these calibrations shall be maintained for two years after the calibration date for inspection by the board.

410 IAC 5-5-7

Indiana State Department of Health; Rule HRH-2,PT E,Sec E.104; filed May 26, 1978, 3:30 pm: 1 IR 200; filed Feb 29, 1984, 10:10 am: 7 IR 929; readopted filed Jul 11, 2001, 2:23 p.m.: 24 IR 4234; readopted filed May 22, 2007, 1:44 p.m.: 20070613-IR-410070141RFA; readopted filed Sep 11, 2013, 3:19 p.m.: 20131009-IR-410130346RFA
Readopted filed 11/13/2019, 3:14 p.m.: 20191211-IR-410190391RFA