Item | Minimum duration of autoclaving at 121° ± 1° C | |
Membrane filters and pads | 10 minutes | |
Carbohydrate-containing media (lauryl tryptose, brilliant green lactose bile broth, etc.) | 12-15 minutes | |
Contaminated materials and discarded tests | 30 minutes | |
Membrane filter assemblies (wrapped), sample collection bottles (empty), and individual glassware items | 15 minutes | |
Rinse water volumes of 500 mL to 1000 mL | 45 minutes | |
Rinse water volumes in excess of 1000 mL | Time adjusted for volume; check for sterility | |
Dilution water blanks | 15 minutes |
Ill. Admin. Code tit. 77, § 465.350
Amended at 35 Ill. Reg. 14494, effective August 12, 2011