Ill. Admin. Code tit. 32 § 340.540

Current through Register Vol. 48, No. 43, October 25, 2024
Section 340.540 - Calibration of Survey Instruments
a) Unless specified in another Part, a licensee shall have each survey instrument used to show compliance with this Part calibrated before first use, annually and following a repair that affects the calibration. A licensee shall:
1) Calibrate all scales with readings up to 10 mSv (1000 mrem) per hour with a radiation source;
2) Calibrate two separated readings on each scale or decade that will be used to show compliance; and
3) Conspicuously note on the instrument the date of calibration.
b) A licensee may not use survey instruments if the difference between the indicated exposure rate and the calculated exposure rate is more than 20 percent.

Ill. Admin. Code tit. 32, § 340.540

Added at 29 Ill. Reg. 20841, effective December 16, 2005