The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:
a) The source-specific input parameters required by the dose calculation algorithm;b) The accuracy of dose, dwell time and treatment time calculations at representative points;c) The accuracy of isodose plots and graphic displays;d) The accuracy of the software used to determine sealed source positions from radiographic images; ande) The accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.Ill. Admin. Code tit. 32, § 335.8230
Added at 30 Ill. Reg. 9029, effective April 28, 2006
Amended at 46 Ill. Reg. 966, effective 12/21/2021