Current through November, 2024
Section 11-45-103 - Inspection and maintenance(a) Each licensee shall ensure that checks are made for obvious defects in x-ray systems before use each day.(b) Each licensee shall conduct a program of at least quarterly inspection and maintenance of x-ray systems to assure proper functioning of components important to safety. All appropriate parts shall be maintained in accordance with manufacturer's specifications. Records of inspection and maintenance shall be maintained for inspection by the department for three years.(c) If any inspection reveals damage to components critical to radiation safety, the x-ray system shall be removed from service and labeled as defective until repairs have been made.(d) Inspections of x-ray systems utilized for nondestructive testing shall be conducted by, or under the supervision of, a qualified medical physicist.[Eff 11/12/99] (Auth: HRS §§ 321-10, 321-11, 321-71) (Imp: HRS §§ 321-1, 321-11(21), 321-71)