Section 19-25d-6 - [Repealed effective 10/1/2025] Special requirements for mobile diagnostic radiographic equipment(a)Equipment(1) All requirements of section 19-25d-5 apply except subdivision (a) (5).(2) The exposure control switch shall be of the dead-man type and shall be so arranged that the operator can stand at least six feet from the patient and well away from the useful beam.(b)Structural shielding When a mobile unit is used routinely in one location, it shall be considered a fixed installation subject to the shielding requirements specified in sections 19-25d-3(c) and 19-25d-5(b).(c)Operating procedures(1) All provisions of subsection 19-25d-5(c) apply except subdivision (2).(2) The target-to-skin distance shall be not less than twelve inches.(3) Personnel monitoring shall be required for all individuals operating mobile x-ray equipment.Conn. Agencies Regs. § 19-25d-6
Effective October 1, 1982; repealed 10/1/2025