"FOR X-RAY FIELD LIMITATION SYSTEM FAILURE"
Stationary radiographic systems shall be required to have the x-ray control, including the exposure switch, permanently mounted in a protected (shielded) area so that the operator is required to remain in that protected area during the entire exposure. Design of the operator protected area shall be consistent with Appendix 6B.
When any one or combination of mobile or portable x-ray systems are:
The following requirements apply for any fixed x-ray tube potential within the range of 40 percent to 100 percent of the maximum rated value:
This is:
Where Xi and X2 are the average mGy/mAs values obtained at each of two consecutive mAs selector settings or at two settings differing by no more than a factor of 2 where the mAs selector provides continuous selection.
For equipment manufactured after May 3, 1994, the average ratios of air kerma to the indicated milliampere-seconds product (mGy/mAs) obtained at any two consecutive mAs selector settings shall not differ by more than 0.10 times their sum.
This is:
Where Xi and X2 are the average mGy/mAs values obtained at each of two consecutive mAs selector settings or at two settings differing by no more than a factor of 2 where the mAs selector provides continuous selection.
Radiation emitted from the x-ray tube shall not exceed:
Except during veterinary field operations where it is impractical to do so, a tube stand or other mechanical support shall be used for portable x-ray systems that are not intended to be handheld during operation.
6 CCR 1007-1-6.6