Appendix 6 CCR 1007-1-2-F - BONE DENSITOMETRY (BD) ADEQUATE RADIATION SAFETY TRAINING AND EXPERIENCE

Each operator of a dual-energy x-ray absorptiometry system used on a living human shall meet the following education and experience requirements:

2F.1 Is certified or registered:

2F.1.1 As R.T.(R), R.T.(M), R.T.(N), R.T.(T), or CNMT; or

2F.1.2 By The International Society for Clinical Densitometry (ISCD), combined with or including the didactic radiation safety training in 2F.2.1.1, 2F.2.1.2 and 2F.2.1.3; or

2F.1.3 By A specialty board determined by the department to have substantially equivalent requirements for certification;

Or

2F.2 Is accepted by the Department as having satisfactorily completed:

2F.2.1 At least 30 hours of didactic training recognized by the Department that provided the minimum hours of instruction (as part of, or in addition to, specialty certificate and equipment operation training) in the specific subjects listed in 2F.2.1.1 through 2F.2.1.9:

RADIATION SAFETY:

2F.2.1.1 Basic X-Ray Physics-2 hours

(1) Structure of matter and the atom

(2) General description of production of x-rays

(3) X-ray emission, quantity and quality

(4) Function of filtration and effects it has on x-ray beam collimation

(5) Types of function of beam limiting devices

(6) Design, features and functions of x-ray tubes

(7) Circuitry of the x-ray machine

2F.2.1.2 Radiobiology-2 hours

(1) Effects of ionizing radiation to the human body

(2) Molecular and cellular radiobiology

(3) Factors that cause somatic and genetic damage

2F.2.1.3 Radiation Protection-5 hours

(1) ALARA

(2) Shielding materials

(3) Radiation quantity and units of measurement

(4) Basic interactions of x-ray with matter

(5) Primary and secondary scatter

(6) Importance of time, distance, shielding

(7) Maximum permissible dose: occupational and public

(8) Patient protection

(a) Patient instruction

(b) Comparison levels of radiation

(i) Natural background radiation

(ii) Central DXA

(iii) Peripheral DXA

2F.2.1.4 Basic Concepts-8 hours

(1) Osteoporosis

(a) World Health Organization definition and diagnostic criteria

(b) Primary vs. secondary

(c) Type I (postmenopausal) vs. Type II (senile)

(d) Risk factors

(i) Controllable (smoking, calcium intake, estrogen, medications)

(ii) Uncontrollable (heredity, race, gender, age, medical conditions)

(2) Bone physiology

(a) Functions of bone

(i) Structural support and protection

(ii) Storage of essential minerals

(b) Types of bone

(i) Cortical

(ii) Trabecular

(c) Bone remodeling cycle

(i) Resorption / formation

(ii) Osteoblasts/osteoclasts

(d) Bone health

(i) Nutrition

(ii) Exercise

(3) BMD testing methods (anatomical sites scanned, key advantages and disadvantages)

(a) Dual-energy X-ray Absorptiometry (DXA)

(b) Single X-ray Absorptiometry (SXA)

(c) Quantitative Ultrasound (QUS)

(d) Radiographic Absorptiometry (RA)

(4) Measuring BMD

(a) Basic statistical concepts

(i) Mean

(ii) Standard deviation

(iii) Coefficient of variation

(b) Reporting patient results

(i) BMD formula

(ii) Z-score

(iii) T-score

2F.2.1.5 Equipment Operation & Quality Control-6 hours

(1) Computer console

(a) Major components

(b) File management

(2) Fundamentals of x-ray energy production

(a) Properties of x-ray beam: quality (kVp), quantity (mA), duration/time (s)

(b) Filters and collimators

(c) X-ray energy production: single; dual

(3) Types of DXA systems

(a) Pencil beam systems

(b) Fan beam systems

(c) Cone beam systems

(4) Quality control

(a) Equipment safety (electrical, pinch points, emergency stop)

(b) Use of phantoms and/or calibration

(c) Troubleshooting

(i) Shift or drift

(ii) Pass / fail

(d) Record maintenance

(5) Determining quality in BMD

(a) Precision (definition)

(b) Accuracy (definition)

(c) Factors affecting accuracy and precision

(i) Scanner

(ii) Operator

(iii) Patient

2F.2.1.6 DXA Scanning of Finger and Heel (OS CALCIS)-1 hour

(1) Anatomy

(a) Regions of interest

(b) Bony landmarks

(c) Radiographic appearance

(2) Scan acquisition

(a) Patient instructions

(b) Patient positioning

(c) Evaluating pre-set scan parameters

(3) Scan analysis: BMD, T score, Z score

(4) Common problems

(a) Nonremovable artifacts

(b) Fractures or pathology

2F.2.1.7 DXA Scanning of Forearm-2 hours

(1) Anatomy

(a) Regions of interest

(b) Bony landmarks

(c) Radiographic appearance

(d) Adjacent structures

(2) Scan acquisition

(a) Patient instructions

(b) Patient positioning

(c) Evaluating pre-set scan parameters

(3) Scan analysis

(a) Accurate ROI placement

(b) BMC, area, and BMD

(c) T-score, Z-score

(4) Common problems

(a) Poor bone edge detection

(b) Nonremovable artifacts

(c) Variant anatomy

(d) Fractures or pathology

(5) Follow-up scans

(a) Unit of comparison: BMD, T-score

(b) Reproduce baseline study

2F.2.1.8 DXA Scanning of Lumbar Spine-2 hours

(1) Anatomy

(a) Regions of interest

(b) Bony landmarks

(c) Radiographic appearance

(d) Adjacent structures

(2) Scan acquisition

(a) Patient instructions

(b) Patient positioning

(c) Evaluating pre-set scan parameters

(3) Scan analysis

(a) Accurate ROI placement

(b) BMC, area, and BMD

(c) T-score, Z-score

(4) Common problems

(a) Poor bone edge detection

(b) Nonremovable artifacts

(c) Variant anatomy

(d) Fractures or pathology

(5) Follow-up scans

(a) Unit of comparison: BMD, T score

(b) Reproduce baseline study

2F.2.1.9 DXA Scanning of Proximal Femur-2 hours

(1) Anatomy

(a) Regions of interest

(b) Bony landmarks

(c) Radiographic appearance

(d) Adjacent structures

(2) Scan acquisition

(a) Patient instructions

(b) Patient positioning

(c) Evaluating pre-set scan parameters

(3) Scan analysis

(a) Accurate ROI placement

(b) BMC, area, and BMD

(c) T-score, Z-score

(4) Common problems

(a) Poor bone edge detection

(b) Nonremovable artifacts

(c) Variant anatomy

(d) Fractures or pathology

(5) Follow-up scans

(a) Unit of comparison: BMD, T-score

(b) Reproduce baseline study;

and

2F.2.2 At least 480 hours of clinical training during which time DXA examinations are performed only under direct supervision of a Colorado qualified bone densitometry equipment operator or other qualified trainer:

2F.2.3 Performance of the following imaging procedures (at least 30 examinations in total, with record of each examination kept on file):

2F.2.3.1 DXA scanning of the forearm-10 examinations;

2F.2.3.2 DXA scanning of the lumbar spine-10 examinations;

2F.2.3.3 DXA scanning of the proximal femur-10 examinations;

and

2F.2.4 A passing score on the American Registry of Radiologic Technologists (ARRT) Bone Densitometry Equipment Operator Examination.

and

2F.2.5 Has maintained a minimum of eighteen (18) hours continuing education every three years, documented by certificate(s) or other attestation(s) of satisfactory completion.

6 CCR 1007-1-2-F

Colorado Register, Vol 37, No. 14. July 25, 2014, effective 8/14/2014
38 CR 02, January 25, 2015, effective 2/14/2015
38 CR 05, March 10, 2015, effective 3/30/2015
38 CR 12, June 25, 2015, effective 7/15/2015
38 CR 14, July 25, 2015, effective 8/14/2015
39 CR 02, January 25, 2016, effective 2/14/2016
39 CR 16, August 25, 2016, effective 9/14/2016
39 CR 22, November 25, 2016, effective 12/15/2016
40 CR 11, June 10, 2017, effective 6/30/2017
40 CR 20, October 25, 2017, effective 11/14/2017
42 CR 24, December 25, 2019, effective 1/14/2020
43 CR 14, July 25, 2020, effective 8/14/2020
43 CR 18, September 25, 2020, effective 10/15/2020
44 CR 11, June 10, 2021, effective 7/15/2021
44 CR 14, July 25, 2021, effective 8/14/2021
45 CR 22, November 25, 2022, effective 12/15/2022
47 CR 06, March 25, 2024, effective 4/14/2024