5 Colo. Code Regs. § 1005-2-4.3

Current through Register Vol. 47, No. 11, June 10, 2024
Section 5 CCR 1005-2-4.3 - Analytic EBAT requirements include:
4.3.1 Providing the subject instruction for delivery of a breath sample that contains end-expiratory air from the lungs.
4.3.2 Starting the test sequence and following the test instructions displayed by the certified EBAT instrument.
4.3.3 Providing a clean mouthpiece with each breath sample provided by the subject.
4.3.4 Observing the subject through completion of the second breath sample to look for signs of belching, regurgitation, or intake of any foreign material into the mouth cavity. If such observations occur, the test sequence must be discontinued by the certified EBAT instructor or operator and another 20-minute deprivation period must be repeated under the same conditions prior to retesting.
4.3.5 Removal of the subject from the area in close proximity to the certified EBAT instrument during the two-minute period between breath samples in order to prevent tampering of the instrument during the test sequence.

5 CCR 1005-2-4.3

41 CR 04, February 25, 2018, effective 3/17/2018
42 CR 02, January 25, 2019, effective 2/14/2019