Section 120.584 - Full Inspection Servicing for Teletherapy and Gamma Stereotactic Radiosurgery Units(A) A licensee shall have each teletherapy unit and gamma stereotactic radiosurgery unit fully inspected and serviced during each source replacement to assure proper functioning of the source exposure mechanism and other safety components. The interval between each full-inspection servicing shall not exceed five years for each teletherapy unit and shall not exceed seven years for each gamma stereotactic radiosurgery unit.(B) This inspection and servicing may only be performed by persons specifically licensed to do so by the Agency, an Agreement State, or the U.S. Nuclear Regulatory Commission.(C) A licensee shall maintain a record of the inspection and servicing in accordance with 105 CMR 120.592(L).Amended by Mass Register Issue 1480, eff. 10/14/2022.