42 C.F.R. § 493.933

Current through May 31, 2024
Section 493.933 - [Effective 7/11/2024] Endocrinology
(a)Program content and frequency of challenge. To be approved for proficiency testing for endocrinology, a program must provide a minimum of five samples per testing event. There must be at least three testing events at approximately equal intervals per year. The annual program must provide samples that cover the clinically relevant range of values that would be expected in patient specimens. The samples may be provided through mailed shipments.
(b)Challenges per testing event. The minimum number of challenges per testing event a program must provide for each analyte or test procedure is five serum, plasma, blood, or urine samples.

Table 1 to Paragraph (b) -Analyte or Test

Cancer antigen (CA) 125.
Carcinoembryonic antigen (CEA).
Cortisol.
Estradiol.
Folate, serum.
Follicle stimulating hormone.
Free thyroxine.
Human chorionic gonadotropin (HCG) (excluding urine pregnancy tests done by visual color comparison categorized as waived tests).
Luteinizing hormone.
Parathyroid hormone.
Progesterone.
Prolactin.
Testosterone.
T3 Uptake.
Triiodothyronine.
Thyroid-stimulating hormone.
Thyroxine.
Vitamin B12.

(c)Evaluation of a laboratory's analyte or test performance. HHS approves only those programs that assess the accuracy of a laboratory's responses in accordance with paragraphs (c)(1) through (5) of this section.
(1) To determine the accuracy of a laboratory's response for qualitative and quantitative endocrinology tests or analytes, a program must compare the laboratory's response for each analyte with the response that reflects agreement of either 80 percent or more of 10 or more referee laboratories or 80 percent or more of all participating laboratories. Both methods must be attempted before the program can choose to not grade a PT sample.
(2) For quantitative endocrinology tests or analytes, the program must determine the correct response for each analyte by the distance of the response from the target value. After the target value has been established for each response, the appropriateness of the response must be determined by using either fixed criteria based on the percentage difference from the target value or the number of standard deviations (SDs) the response differs from the target value.

Table 2 to Paragraph (c)(2)-Criteria for Acceptable Performance

The criteria for acceptable performance are- Analyte or testCriteria for acceptable performance
Cancer antigen (CA) 125Target value ±20%.
Carcinoembryonic antigen (CEA)Target value ±15% or ±1 ng/mL (greater).
CortisolTarget value ±20%.
EstradiolTarget value ±30%.
Folate, serumTarget value ±30% or ±1 ng/mL (greater).
Follicle stimulating hormoneTarget value ±18% or ±2 IU/L (greater).
Free thyroxineTarget value or ±15% or ±0.3 ng/dL (greater).
Human chorionicTarget value ±18% or ±3
gonadotropin (excluding urine pregnancy tests done by visual color comparison categorized as waived tests)mIU/mL (greater) or positive or negative.
Luteinizing hormoneTarget value ±20%.
Parathyroid hormoneTarget value ±30%.
ProgesteroneTarget value ±25%.
ProlactinTarget value ±20%.
TestosteroneTarget value ±30% or ±20 ng/dL (greater).
T3 uptakeTarget value ±18%.
TriiodothyronineTarget value ±30%.
Thyroid-stimulating hormoneTarget value ±20% or ±0.2 mIU/L (greater).
ThyroxineTarget value ±20% or ±1.0 mcg/dL (greater).
Vitamin B12Target value ±25% or ±30 pg/mL (greater).

(3) The criterion for acceptable performance for qualitative endocrinology tests is positive or negative.
(4) To determine the analyte testing event score, the number of acceptable analyte responses must be averaged using the following formula:

View Image

(5) To determine the overall testing event score, the number of correct responses for all analytes must be averaged using the following formula:

View Image

42 C.F.R. §493.933

87 FR 41239, 8/10/2022; 87 FR 68912, 7/11/2024