The measurement instrument calibration and verification requirements in 40 CFR part 1065, subpart D, apply with the following exceptions:
Table 1 of § 1066.130 -Summary of Required Calibrations and Verifications
Type of calibration or verification | Minimum frequencya |
40 CFR 1065.307 : Linearity verification | The linearity verifications from 40 CFR part 1065 do not apply under this part for engine speed, torque, fuel rate, or intake air flow; the linearity verification described in § 1066.135 applies for the following measurements: Dynamometer speed: See § 1066.220 . Dynamometer torque: See § 1066.220 . |
40 CFR 1065.310 : Torque | This calibration does not apply for testing under this part; see § 1066.220 . |
40 CFR 1065.320 : Fuel flow | This calibration does not apply for testing under this part. |
40 CFR 1065.325 : Intake flow | This calibration does not apply for testing under this part. |
40 CFR 1065.340 : CVS calibration | This calibration does not apply for CVS flow meters calibrated volumetrically as described in § 1066.140 . |
40 CFR 1065.345 : Vacuum leak | Required upon initial installation of the sampling system; recommended within 35 days before the start of an emissions test and after maintenance such as pre-filter changes. |
40 CFR 1065.350(c) , 1065.355(c) , 1065.370(c) , and 1065.375(c) | These provisions do not apply for testing under this part; see § 1066.150 . |
40 CFR 1065.395 : Inertial PM balance and weighing | These verifications do not apply for testing under this part. |
a Perform calibrations and verifications more frequently if needed to conform to the measurement system manufacturer's instructions and good engineering judgment.
40 C.F.R. §1066.130