40 C.F.R. § 1065.790

Current through March 31, 2024
Section 1065.790 - Mass standards
(a)PM balance calibration weights. Use PM balance calibration weights that are certified as NIST-traceable within ±0.1% uncertainty. Make sure your highest calibration weight has no more than ten times the mass of an unused PM-sample medium.
(b)Dynamometer, fuel mass scale, and DEF mass scale calibration weights. Use dynamometer and mass scale calibration weights that are certified as NIST-traceable within ±0.1% uncertainty.

40 C.F.R. §1065.790

88 FR 4687, Jan. 24, 2023
86 FR 34574, 7/29/2021; 88 FR 4687, 3/27/2023