40 C.F.R. § 1065.290

Current through May 31, 2024
Section 1065.290 - PM gravimetric balance
(a)Application. Use a balance to weigh net PM on a sample medium for laboratory testing.
(b)Component requirements. We recommend that you use a balance that meets the specifications in Table 1 of § 1065.205 . Note that your balance-based system must meet the linearity verification in § 1065.307 . If the balance uses internal calibration weights for routine spanning and the weights do not meet the specifications in § 1065.790 , the weights must be verified independently with external calibration weights meeting the requirements of § 1065.790 . While you may also use an inertial balance to measure PM, as described in § 1065.295 , use a reference procedure based on a gravimetric balance for comparison with any proposed alternate measurement procedure under § 1065.10 .
(c)Pan design. We recommend that you use a balance pan designed to minimize corner loading of the balance, as follows:
(1) Use a pan that centers the PM sample media (such as a filter) on the weighing pan. For example, use a pan in the shape of a cross that has upswept tips that center the PM sample media on the pan.
(2) Use a pan that positions the PM sample as low as possible.
(d)Balance configuration. Configure the balance for optimum settling time and stability at your location.

40 C.F.R. §1065.290

73 FR 37300, June 30, 2008, as amended at 75 FR 68462, Nov. 8, 2010