Test day | Line voltage,1 rms | Room temperature,2°C | Comments |
0 | 115 | 25 | Initial set-up and adjustments. |
1 | 125 | 20 | |
2 | 105 | 20 | |
3 | 125 | 30 | Adjustments and/or periodic maintenance permitted at end of tests. |
4 | 105 | 30 | |
5 | 125 | 20 | |
6 | 105 | 20 | Adjustments and/or periodic maintenance permitted at end of tests. |
7 | 125 | 30 | Examine test results to ascertain if further testing is required. |
8 | 105 | 30 | |
9 | 125 | 20 | Adjustments and/or periodic maintenance permitted at end of tests. |
10 | 105 | 20 | |
11 | 125 | 30 | |
12 | 105 | 30 | Adjustments and/or periodic maintenance permitted at end of tests. |
13 | 125 | 20 | |
14 | 105 | 20 | |
15 | 125 | 30 |
1 Voltage specified shall be controlled to ±1 volt.
2 Temperatures shall be controlled to ±1 °C.
40 C.F.R. 53 app Table B-4 to Subpart B of Part 53