Current through November 30, 2024
Section 878.4783 - Negative pressure wound therapy device for reduction of wound complications(a)Identification. A negative pressure wound therapy device for reduction of wound complications is a powered suction pump intended for wound management and reduction of wound complications via application of negative pressure to the wound, which removes fluids, including wound exudate, irrigation fluids, and infectious materials. This device type is intended for use with wound dressings classified under § 878.4780 . This classification does not include devices intended for organ space wounds.(b)Classification. Class II (special controls). The special controls for this device are:(1) Clinical data must demonstrate that the device performs as intended under anticipated conditions of use and evaluate the following:(i) Wound complication rates; and(2) The patient-contacting components of the device must be demonstrated to be biocompatible.(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the labeled shelf life.(5) Usability testing must demonstrate that intended users can correctly use the device, based solely on reading the instructions for use.(6) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested in a worst-case scenario for the intended use life:(i) Ability to maintain pressure levels at the wound site under a worst-case scenario for the intended use life;(ii) Fluid removal rate consistent with the wound types specified in the indications for use; and(iii) Timely triggering of all alarms.(7) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of the device.(8) Software verification, validation, and hazard analysis must be performed.(9) Labeling must include the following:(i) Instructions for use;(ii) A summary of the device technical specifications, including pressure settings, modes (e.g., continuous or intermittent), alarms, and safety features;(iii) Compatible components and devices;(iv) A summary of the clinical evidence for the indications for use;(v) A shelf life for sterile components; and(vi) Use life and intended use environments.(10) For devices intended for use outside of a healthcare facility, patient labeling must include the following:(i) Information on how to operate the device and its components and the typical course of treatment;(ii) Information on when to contact a healthcare professional; and 86 FR 70734 , Dec. 10, 2021