Instrumental Parameters for Flame AAS Analysis
Atomic Absorption Spectophotometer (Perkin-Elmer Model 603)
Flame: | Air/Acetylene lean, blue | |
Oxidant Flow: | 55 | |
Fuel Flow: | 32 | |
Wavelength: | 228.8 nm | |
Slit: | 4 (0.7 nm) | |
Range: | UV | |
Signal: | Concentration (4 exp) | |
Integration Time: | 3 sec |
Cal. Code Regs. Tit. 8, div. 1, ch. 4, subch. 7, grp. 16, art. 110, app E