Appendix B - Start up and Shut down Procedure (Non-Mandatory)

(a) Start up.

(1) Turn off ultraviolet sterilizer if so equipped.

(2) Turn on all blowers and cabinet illumination lights.

(3) Allow 5 minutes of operation to purge system, check flow alarm system audio and visual alarm function if so equipped.

(4) Decontaminate readily accessible interior surfaces with a disinfectant appropriate for the agents or suspected agents present.

(b) Shut down.

(1) Decontaminate and remove all items from interior work area.

(2) Decontaminate readily accessible interior surfaces with a disinfectant appropriate for the agents or suspected agents present.

(3) Turn on ultraviolet sterilizer if so equipped.

(4) Allow 5 minutes of operation to purge system.

(5) Turn off cabinet blower.

Cal. Code Regs. Tit. 8, div. 1, ch. 4, subch. 7, grp. 16, art. 107 app B

1. New Appendix B filed 10-11-94; operative 11-10-94 (Register 94, No. 41).

Note: Authority cited: Section 142.3, Labor Code. Reference: Section 142.3, Labor Code.