Cal. Code Regs. tit. 17 § 95154

Current through Register 2024 Notice Reg. No. 25, June 21, 2024
Section 95154 - Monitoring and QA/QC Requirements

The GHG emissions data for petroleum and natural gas emissions sources must be quality assured as applicable and as specified in this section. Offshore petroleum and natural gas production facilities must adhere to the monitoring and QA/QC requirements as set forth in 30 CFR Part 250 (July 1, 2011), which is hereby incorporated by reference.

(a) Facility operators must use any of the methods described as follows in this paragraph to conduct leak detection(s) of equipment leaks and through-valve leakage from all source types listed in sections 95153(i), (m), (n) and (o) that occur during a calendar year, except as provided in paragraph (a)(4) of this section.
(1)Optical gas imaging instrument. Use an optical gas imaging instrument for equipment leak detection in accordance with 40 CFR Part 60 , subarticle A, § 60.18 of the Alternative work practice for monitoring equipment leaks, § 60.18(i)(1)(i); § 60.18(i)(2)(i) except that the monitoring frequency shall be annual using the detection sensitivity level of 60 grams per hour as stated in 40 CFR Part 60 , subarticle A, Table 1: Detection Sensitivity Levels; § 60.18(i)(2)(ii) and (iii) except the gas chosen shall be methane, and § 60.18(i)(2)(iv) and (v); § 60.18(i)(3); § 60.18(i)(4)(i) and (v); including the requirements for daily instrument checks and distances, and excluding requirements for video records (July 1, 2011, which is hereby incorporated by reference). Any emissions detected by the optical gas imaging instrument is a leak unless screened with Method 21 ( 40 CFR Part 60 , appendix A-7 (July 1, 2011), which is hereby incorporated by reference) monitoring, in which case 10,000 ppm or greater is designated a leak. In addition, facility operators must operate the optical gas imaging instrument to image the source types required by this subarticle in accordance with the instrument manufacturer's operating parameters. Unless using methods in paragraph (a)(2) of this section, an optical gas imaging instrument must be used for all source types that are inaccessible and cannot be monitored without elevating the monitoring personnel more than two meters above a support surface.
(2)Method 21. Use the equipment leak detection methods in 40 CFR Part 60 , appendix A-7, Method 21 (July 1, 2011). If using Method 21 monitoring, if an instrument reading of 10,000 ppm or greater is measured, a leak is detected. Inaccessible emissions sources, as defined in 40 CFR Part 60 , are not exempt from this subarticle. Owners or operators must use alternative leak detection devices as described in paragraph (a)(1) or (a)(2) of this section to monitor inaccessible equipment leaks or vented emissions.
(3)Infrared laser beam illuminated instrument. Use an infrared laser beam illuminated instrument for equipment leak detection. Any emissions detected by the infrared laser beam illuminated instrument is a leak unless screened with Method 21 monitoring, in which case 10,000 ppm or greater is designated a leak. In addition, the facility operator must operate the infrared laser beam illuminated instrument to detect the source types required by this subarticle in accordance with the instrument manufacturer's operating instructions.
(4)Optical gas imaging instrument. An optical gas imaging instrument must be used for all source types that are inaccessible and cannot be monitored without elevating the monitoring personnel more than 2 meters above a support surface.
(5)Acoustic leak detection device. Use the acoustic leak detection device to detect through-valve leakage. When using the acoustic leak detection device to quantify the through-valve leakage, use the instrument manufacturer's calculation methods to quantify the through-valve leak. When using the acoustic leak detection device, if a leak of 3.1 scf per hour or greater is calculated, a leak is detected. In addition, the facility operator must operate the acoustic leak detection device to monitor the source valves required by this subarticle in accordance with the instrument manufacturer's operating parameters. Acoustic stethoscope type devices designed to detect through valve leakage when put in contact with the valve body and that provide an audible leak signal but do not calculate a leak rate can be used to identify non-leakers with subsequent measurement required to calculate the rate if through-valve leakage is identified. Leaks are reported if a leak rate of 3.1 scf per hour or greater is measured. In addition, the facility operator must operate the acoustic leak detection device to monitor the source valves required by this subarticle in accordance with the instrument manufacturer's operating parameters.
(b) The operator must operate and calibrate all flow meters, composition analyzers and pressure gauges used to measure quantities reported in section 95153 according to the procedures in section 95103(k) and the procedures in paragraph (b) of this section. Pursuant to section 95109 of this article, the facility operator may use an appropriate standard method published by a consensus-based standards organization if such a method exists or use an industry standard practice.
(c) Use calibrated bags (also known as vent bags) only where the emissions are at near-atmospheric pressures and below the maximum temperature specified by the vent bag manufacturer such that the vent bag is safe to handle. The bag opening must be of sufficient size that the entire emission can be tightly encompassed for measurement till the bag is completely filled.
(1) Hold the bag in place enclosing the emissions source to capture the entire emissions and record the time required for completely filling the bag. If the bag inflates in less than one second, assume one second inflation time.
(2) Perform three measurements of the time required to fill the bag, report the emissions as the average of the three readings.
(3) Estimate natural gas volumetric emissions at standard conditions using calculations in section 95153(r).
(4) Estimate CH4 and CO2 volumetric and mass emissions from volumetric natural gas emissions using the calculations in sections 95153(s) and (t).
(d) Use a high volume sampler to measure emissions within the capacity of the instrument.
(1) A technician following manufacturer instructions shall conduct measurements, including equipment manufacturer's operating procedures and measurement methodologies relevant to using a high volume sampler, including positioning the instrument for complete capture of the equipment leak without creating backpressure on the source.
(2) If the high volume sampler, along with all attachments available from the manufacturer, is not able to capture all the emissions from the source then use anti-static wraps or other aids to capture all emissions without violating operating requirements as provided in the instrument manufacturer's manual.
(3) Estimate natural gas volumetric emissions at standard conditions using calculations in section 95153(r). Estimate CH4 and CO2 volumetric and mass emissions from volumetric natural gas emissions using the calculations in sections 95153(s) and (t).
(4) Calibrate the instrument at 2.5 percent methane with 97.5 percent air and 100 percent CH4 by using calibrated gas samples by following manufacturer's instructions for calibration.
(e) Peng-Robinson Equation of State means the equation of state defined by Equation 38 of this section.

Click here to view image

Eq. 38)

Where:

p = Absolute pressure.

R = Universal gas constant

T = Absolute temperature.

Vm = Molar volume.

Click here to view image

Where:

[OMEGA] = Acentric factor of the species.

Tc = Critical temperature.

Pc = Critical pressure.

Cal. Code Regs. Tit. 17, § 95154

1. New section filed 12-14-2011; operative 1-1-2012 pursuant to Government Code section 11343.4 (Register 2011, No. 50).
2. Repealer and new section filed 12-19-2012; operative 1-1-2013 pursuant to Government Code section 11343.4 (Register 2012, No. 51).
3. Amendment of subsection (a) and repealer of subsections (f)-(f)(4)(D) filed 12-31-2013; operative 1-1-2014 pursuant to Government Code section 11343.4(b)(3) (Register 2014, No. 1).

Note: Authority cited: Sections 38510, 38530, 39600, 39601, 39607, 39607.4 and 41511, Health and Safety Code. Reference: Sections 38530, 39600 and 41511, Health and Safety Code.

1. New section filed 12-14-2011; operative 1-1-2012 pursuant to Government Code section 11343.4 (Register 2011, No. 50).
2. Repealer and new section filed 12-19-2012; operative 1-1-2013 pursuant to Government Code section 11343.4 (Register 2012, No. 51).
3. Amendment of subsection (a) and repealer of subsections (f)-(f)(4)(D) filed 12-31-2013; operative 1-1-2014 pursuant to Government Code section 11343.4(b)(3) (Register 2014, No. 1).