(a) Each radiographic exposure device shall be provided with a lock or outer locked container designed to prevent unauthorized or accidental exposure and shall be kept locked, and if a keyed-lock, with the key removed, at all times except during authorized use or when under the direct surveillance of a radiographer or radiographer's assistant. In addition, during radiographic operations a sealed source assembly shall be secured in the shielded position each time the source is returned to that position.(b) Each storage container and source changer shall be provided with a lock and kept locked, and if a keyed-lock, with the key removed, when containing a sealed source except when the container is under the direct surveillance of a radiographer or radiographer's assistant.(c) Locked radiographic exposure devices, storage containers and source changers shall be physically secured to prevent tampering or unauthorized removal.Cal. Code Regs. Tit. 17, § 30332.1
1. Renumbering and amendment of former Section 30332(b) to Section 30332.1 filed 7-7-86; effective thirtieth day thereafter (Register 86, No. 28). For prior history, see Registers 72, No. 42 and 67, No. 46.
2. Amendment of section and NOTE filed 4-11-2008; operative 5-11-2008 (Register 2008, No. 15). Note: Authority cited: Sections 114975, 115000, 131050, 131051 and 131200, Health and Safety Code. Reference: Sections 114965, 114970, 115000, 115060, 115230 and 115235, Health and Safety Code.