Cal. Code Regs. tit. 17 § 30316.10

Current through Register 2025 Notice Reg. No. 2, January 10, 2025
Section 30316.10 - Evaluations of New, Reassembled, Replaced or Repaired Equipment
(a) Whenever the following occur, the affected equipment shall meet the requirements specified in sections 30316 and 30316.20 before the affected equipment is used to produce or process mammograms:
(1) Installation of a new radiation machine or processor;
(2) Disassembly and reassembling of a radiation machine or processor at the same or a new location; or
(3) Repair or replacement of any major component of the mammography system.
(b) The evaluation to determine if equipment specified in subsections (a)(1) through (3) are in compliance with subsection (a) shall be performed by a mammography medical physicist or an individual under the direct supervision of a mammography medical physicist.
(c) Documentation of the tests performed, the analysis of data obtained, corrective actions, and the effectiveness of those actions taken pursuant to this section shall be maintained in accordance with section 30317.20.

Cal. Code Regs. Tit. 17, § 30316.10

1. New section filed 2-10-2003 as an emergency; operative 2-10-2003 (Register 2003, No. 7). A Certificate of Compliance must be transmitted to OAL by 6-10-2003 or emergency language will be repealed by operation of law on the following day.
2. Certificate of Compliance as to 2-10-2003 order, including amendment of subsections (a), (c)(1)-(2) and (d)(1)-(2), transmitted to OAL 6-6-2003 and filed 7-18-2003 (Register 2003, No. 29).
3. Amendment of section heading, section and NOTE filed 12-18-2019; operative 7-1-2020 pursuant to Government Code section 11343.4(b)(2) (Register 2019, No. 51).

Note: Authority cited: Sections 114975, 115000, 115060 and 131200, Health and Safety Code. Reference: Sections 115060, 115115, 131050 and 131051, Health and Safety Code.

1. New section filed 2-10-2003 as an emergency; operative 2-10-2003 (Register 2003, No. 7). A Certificate of Compliance must be transmitted to OAL by 6-10-2003 or emergency language will be repealed by operation of law on the following day.
2. Certificate of Compliance as to 2-10-2003 order, including amendment of subsections (a), (c)(1)-(2) and (d)(1)-(2), transmitted to OAL 6-6-2003 and filed 7-18-2003 (Register 2003, No. 29).
3. Amendment of section heading, section and Note filed 12-18-2019; operative 7/1/2020 pursuant to Government Code section 11343.4(b)(2) (Register 2019, No. 51).