Ariz. Admin. Code § tit. 13, ch. 10, art. 1, exh. G-2

Current through Register Vol. 30, No. 24, June 14, 2024
Exhibit G-2 - Standard Calibration Check Procedure, Intoxilyzer Model 8000

THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A)

ARIZONA DEPARTMENT OF PUBLIC SAFETY

STANDARD QUALITY ASSURANCE PROCEDURES

INTOXILYZER MODEL 8000

STANDARD CALIBRATION CHECK PROCEDURE

QA SPECIALIST ________________________________________ AGENCY ____________________

DATE _______________________________________TIME _________________________________

INTOXILYZER SERIAL # ______________________________LOCATION _______________________

( ) 1. Ensure that gas tank is attached to instrument and contains a standard alcohol concentration solution _________AC.

OR

Pour a standard alcohol concentration solution _________AC, into a clean dry simulator and assemble the simulator. Ensure that a tight seal has been made. Turn on the simulator and allow temperature to reach 34° C ± 0.2° C

( ) 2. Intoxilyzer 8000 display reads "PUSH BUTTON TO START"

( ) 3. Go to the "Control Testing Menu". Select "D" for dry control test or "W" for wet control test. After selection is made press ENTER.

( ) 4. Air blank completed.

( ) 5. Calibration check completed. Test results 0._____________AC.

( ) 6. Air blank completed.

( ) 7. Remove printed record. Attach the record to the completed checklist.

SIGNATURE _________________________________________________________________________

DPS Form Exh G-2 (Rev 05-01)

Ariz. Admin. Code tit. 13, ch. 10, art. 1, exh. G-2

New Exhibit G-2 made by final rulemaking at 12 A.A.R. 1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).