Phil LangstonDownload PDFPatent Trials and Appeals BoardSep 26, 201914971504 - (D) (P.T.A.B. Sep. 26, 2019) Copy Citation UNITED STATES PATENT AND TRADEMARK OFFICE UNITED STATES DEPARTMENT OF COMMERCE United States Patent and Trademark Office Address: COMMISSIONER FOR PATENTS P.O. Box 1450 Alexandria, Virginia 22313-1450 www.uspto.gov APPLICATION NO. FILING DATE FIRST NAMED INVENTOR ATTORNEY DOCKET NO. CONFIRMATION NO. 14/971,504 12/16/2015 Phil Langston 2005.63US04 8828 24113 7590 09/26/2019 PATTERSON THUENTE PEDERSEN, P.A. 80 SOUTH 8TH STREET 4800 IDS CENTER MINNEAPOLIS, MN 55402-2100 EXAMINER TOTH, KAREN E ART UNIT PAPER NUMBER 3791 MAIL DATE DELIVERY MODE 09/26/2019 PAPER Please find below and/or attached an Office communication concerning this application or proceeding. The time period for reply, if any, is set in the attached communication. PTOL-90A (Rev. 04/07) UNITED STATES PATENT AND TRADEMARK OFFICE ____________ BEFORE THE PATENT TRIAL AND APPEAL BOARD ____________ Ex parte PHIL LANGSTON Appeal 2019-002778 Application 14/971,504 Technology Center 3700 ____________ Before JOHN C. KERINS, MICHAEL J. FITZPATRICK, and MICHELLE R. OSINSKI, Administrative Patent Judges. FITZPATRICK, Administrative Patent Judge. DECISION ON APPEAL Appellant, Phil Langston,1 appeals under 35 U.S.C. § 134(a) from the Examiner’s final decision rejecting claims 23–43 and 45–53. We have jurisdiction under 35 U.S.C. § 6(b). We affirm in part. 1 Appellant states that “Teleflex Incorporated and its affiliates” are additional real parties in interest by virtue of Teleflex Incorporated being an exclusive licensee of the instant Application. Appeal Br. 2. Appeal 2019-002778 Application 14/971,504 2 STATEMENT OF THE CASE The Specification The Specification’s disclosure relates generally to the field of cardiac catheters and, more specifically, to cardiac catheters for performing procedures for the testing of aortic stenosis, a condition in which the aortic valve has become narrowed and does not open normally. Spec. 1:11–17. The Rejected Claims Claims 23–43 and 45–53 are rejected; no other claims are pending. Final Act. 1. Independent claims 23 and 46 are illustrative and reproduced below. 23. A catheter comprising: a manifold portion including a first connector and a second connector; and a catheter body including, a proximal dual lumen portion including exactly two longitudinally-extending lumens, an inner lumen and a surrounding outer lumen, that share a common axis, the outer lumen surrounded by a proximal outer wall and the inner lumen surrounded by a proximal inner wall that is structured to withstand high-pressure injections; and a distal single lumen portion including an extension of the inner lumen, the single lumen portion surrounded by a distal outer wall, the first connector in fluid communication with the inner lumen and the second connector in fluid communication with the outer lumen. 46. A catheter comprising: a manifold portion including a first connector and a second connector; and a catheter body including, a proximal dual lumen portion including exactly an inner lumen and a Appeal 2019-002778 Application 14/971,504 3 surrounding outer lumen that share a common axis and at least two side holes, the inner lumen in fluid communication with the first connector, the outer lumen in fluid communication with the second connector, and the at least two side holes consecutively aligned with each other along a line that is parallel with the common axis; and a distal single lumen portion forming an extension of the inner lumen, the single lumen portion including a bend, a plurality of side holes, and a pigtail tip. Appeal Br. 29, 33. The Examiner’s Rejections The following rejections, all under 35 U.S.C. § 103(a), are before us for review: 1. claims 23–28 over Martin2 and Bodicky3 (Final Act. 2); 2. claim 29 over Martin, Bodicky, and Andrews4 (id. at 3); 3. claims 30–37 over Martin, Bodicky, and Schweich5 (id. at 4); 4. claims 38–43 over Martin, Bodicky, and Ladika6 (id. at 7); 5. claim 45 over Martin, Bodicky, and Duffy7 (id. at 8); 6. claims 46–49 and 51 over Martin ’380,8 Ladika, and Schweich (id. at 9); 7. claim 52 over Martin ’380, Ladika, Schweich, and Duffy (id. at 12); 2 US 5,976,103, issued Nov. 2, 1999 (“Martin”). 3 US 4,961,731, issued Oct. 9, 1990 (“Bodicky”). 4 US 5,865,721, issued Feb. 2, 1999 (“Andrews”). 5 US 5,782,797, issued July 21, 1998 (“Schweich”). 6 US 4,747,840, issued May 31, 1988 (“Ladika”). 7 US 6,048,332, issued Apr. 11, 2000 (“Duffy”). 8 US 5,480,380, issued Jan. 2, 1996 (“Martin ’380”). Appeal 2019-002778 Application 14/971,504 4 8. claims 46, 47, and 51 over Aigner9 and Ladika (id.); 9. claim 50 over Aigner, Ladika, and Bodicky (id. at 14); and 10. claim 53 over Martin, Bodicky, and Miller (id. at 14). DISCUSSION Rejections 8 and 9—Aigner as Primary Reference Appellant does not argue against the rejection of claims 46, 47, and 51 over Aigner and Ladika or the rejection of claim 50 over Aigner, Ladika, and Bodicky. Thus, Appellant fails to apprise of error in those rejections. Rejections 8 and 9 are affirmed. Rejections 1–5 and 10—Martin as Primary Reference The Examiner found that Martin teaches all of the subject matter of independent claim 23 except for the proximal inner wall of the inner lumen being “structured to withstand high-pressure injections,” as claimed. Final Act. 2–3 (citing Martin 3:13–27, 4:48–61, Figs. 1 and 2). The Examiner found that “Bodicky teaches forming a catheter configured to withstand high pressure injections at pressures up to 1200 pounds per square inch in order to inject a diagnostic fluid.”10 Id. at 3 (citing Bodicky 3:15–20). The Examiner concluded: “It would have been obvious to one of ordinary skill in the art at the time the invention was made to have made the catheter of Martin and made [sic, make] it suitable for high pressure injections, as taught by Bodicky, in order to allow its use in diagnostics.” Id. Appellant argues that this reason for why a person of ordinary skill in the art allegedly would have so modified Martin’s catheter in view of 9 US 4,666,426, issued May 19, 1987 (“Aigner”). 10 The cited portion of Bodicky actually recites “1000 pounds per square inch” not the 1,200 asserted by the Examiner. Bodicky 3:19–20. Appeal 2019-002778 Application 14/971,504 5 Bodicky is insufficient. Appeal Br. 25–26; see also KSR Int’l Co. v. Teleflex Inc., 550 U.S. 398, 418 (2007) (“requiring an apparent reason to combine the known elements in the fashion claimed by the patent at issue”); InTouch Techs., Inc. v. VGO Commc’ns, Inc., 751 F.3d 1327, 1347 (Fed. Cir. 2014) (Obviousness requires “evidence that a skilled artisan would have been motivated to combine the teachings of the prior art references to achieve the claimed invention, and that the skilled artisan would have had a reasonable expectation of success in doing so.”). We agree with Appellant. Martin teaches “dual lumen catheters for use in hemodialysis treatments and more particularly to a dual lumen catheter having coaxial intake and return lumens.” Martin 1:13–16. Dual lumen hemodialysis catheters are preferred because they use a single incision rather than two. Id. at 1:26–28. Prior art needles, as well as Martin’s “tip section,” are inserted into veins, not arteries. See id. at 1:21–22 (“The earliest treatments were conducted using two needles in the same vein.”), 2:40–44 (“[I]t is desirable that the tip section be sufficiently soft and pliable to permit this section to take up the local shape of a vein containing the catheter, thereby to avoid applying distorting forces to the vein and to permit prolonged access with a suitable selection of materials.”). Martin does not disclose that cleaned blood is returned to a patient’s vein via high pressure injections, nor would that be expected to be safe or desirable. Bodicky “relates generally to angiographic catheters for injecting a radiopaque material or contrast medium into arteries and/or the heart, for radiological diagnostic purposes.” Bodicky 1:7–10. Tubular portion 20 is constructed of a material “selected to give the catheter the desired torque and column strength necessary to push and rotate the catheter while inserting and Appeal 2019-002778 Application 14/971,504 6 placing the catheter in position in the patient’s heart or coronary artery or other blood vessel.” Id. at 3:7–17. The material is “sufficiently strong to withstand high-pressure fluid injection at pressures on the order of 1000 pounds per square inch.” Id. at 3:17–20. The Examiner proposes that a person of ordinary skill in the art would want to make “the catheter of Martin . . . suitable for high pressure injections, as taught by Bodicky, in order to allow its use in diagnostics.” Final Act. 3. But this begs the question: Why would a person of ordinary skill in the art want to convert Martin’s dual lumen catheter, which is used for hemodialysis, into a diagnostics tool capable of providing high pressure fluid injection? No reason is provided by the Examiner. On the record presented, there is not an apparent reason a person of ordinary skill in the art would have modified Martin in view of Bodicky. The lack of an apparent reason is fatal to all of Rejections 1–5 and 10. For the foregoing reasons, we reverse Rejections 1–5 and 10. Rejections 6 and 7—Martin ’380 as Primary Reference The Examiner found that Martin ’380 teaches all of the subject matter of independent claim 46 except for the distal single lumen portion which forms an extension of the inner lumen “including a bend, a plurality of side holes, and a pigtail tip,” as claimed. Final Act. 9–10 (citing Martin ’380 Figs. 1, 2, 6, and 8). The Examiner found that “Ladika teaches a catheter comprising a single lumen portion with a straight section (element 26), a bend (element 20), and a pigtail portion (element 18) distal to the straight portion (see figure 1), in order to prevent the catheter from puncturing the patient’s tissue. Id. at 10 (citing Ladika 1:60–66). The Examiner concluded: “It would have been obvious to one of ordinary skill in the art at the time the Appeal 2019-002778 Application 14/971,504 7 invention was made to have made the device of Martin[ ]’380 with a distal bend and pigtail section, as taught by Ladika, in order to prevent the catheter from puncturing the patient’s tissue.” Id. Appellant again argues insufficient rationale for combining the cited prior art features. Appeal Br. 27. And again we agree with Appellant. Like Martin, Martin ’380 “relates to dual lumen catheters for use in haemodialysis treatments and more particularly to a dual lumen catheter having intake and return lumens.” Martin ’380 1:5–7. The distal end of the catheter, tip portion 65, is narrow so as to be inserted into, and positioned within, a vein. Id. at Fig. 6, 4:2 (“tip portion 65”), 1:14–15 (“The earliest treatments were conducted using two needles in the same vein.”), 1:38–40 (“As a result, no matter where the catheter may rest against a vein, some of the intake 40 openings remain patent.”). Ladika relates “to heart catheters for use in pulmonary arteriography.” Ladika 1:7–8. The widely accepted technique for pulmonary arteriography (as it is known) is via the femoral approach wherein a catheter is first inserted into the patient’s body through a vein in a lower limb and fed through the inferior vena cavae, into the right atrium of the heart, then through the tricuspid valve and into the right ventricle, then through the pulmonary semilunar valve and into the main or right or left pulmonary artery. Thereafter, the radiopaque fluid is injected through the catheter for disbursement within the selected portion of the pulmonary artery. Id. at 1:21–31. Pigtail 18 on distal end of Ladika’s catheter is used to facilitate complex maneuvering of the catheter during a femoral approach. Id. at 2:4–5. The Examiner proposes that a person of ordinary skill in the art would want to make “the device of Martin ’380 with a distal bend and pigtail Appeal 2019-002778 Application 14/971,504 8 section, as taught by Ladika, in order to prevent the catheter from puncturing the patient’s tissue.” Final Act. 10. But a hemodialysis catheter is inserted merely into a vein. Martin ’380 1:14–15, 1:38–40. It is not advanced into the heart or pulmonary tissue as is Ladika’s pulmonary arteriography catheter. On the record presented, there is not an apparent reason a person of ordinary skill in the art would have modified Martin ’380 in view of Ladika. Further, we agree with Appellant that incorporating Ladika’s distal bend and pigtail into Martin ’380 would render it “unsuitable for its intended placement within a vein and accommodating dialysis of the blood.” Appeal Br. 22. For the foregoing reasons, we reverse Rejection 6, as well as Rejection 7. Appeal 2019-002778 Application 14/971,504 9 SUMMARY Claims Rejected Basis Affirmed Reversed 23–28 §103(a); Martin, Bodicky 23–28 29 §103(a); Martin, Bodicky, Andrews 29 30–37 §103(a); Martin, Bodicky, and Schweich 30–37 38–43 §103(a); Martin, Bodicky, and Ladika 38–43 45 §103(a); Martin, Bodicky, and Duffy 45 46–49, 51 §103(a); Martin ’380, Ladika, Schweich 46–49, 51 52 §103(a); Martin ’380, Ladika, Schweich, Duffy 52 46, 47, 51 §103(a); Aigner, Ladika 46, 47, 51 50 §103(a); Aigner, Ladika, Bodicky 50 53 §103(a); Martin, Bodicky, Miller 53 Overall Outcome 46, 47, 50, 51 23–43, 45, 48, 52 No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.F.R. § 1.136(a)(1)(iv). AFFIRMED-IN-PART Copy with citationCopy as parenthetical citation