Ex Parte BonniDownload PDFBoard of Patent Appeals and InterferencesAug 20, 200810703287 (B.P.A.I. Aug. 20, 2008) Copy Citation UNITED STATES PATENT AND TRADEMARK OFFICE __________ BEFORE THE BOARD OF PATENT APPEALS AND INTERFERENCES __________ Ex parte ARAM BONNI __________ Appeal 2008-33171 Application 10/703,287 Technology Center 3700 __________ Decided: August 20, 2008 __________ Before DEMETRA J. MILLS, ERIC GRIMES, and FRANCISCO C. PRATS, Administrative Patent Judges. PRATS, Administrative Patent Judge. DECISION ON APPEAL This is an appeal under 35 U.S.C. § 134 involving claims to a patient-adjustable urinary incontinence device. The Examiner has rejected the claims as anticipated and obvious. We have jurisdiction under 35 U.S.C. § 6(b). We affirm-in-part. 1 Heard August 12, 2008. Appeal 2008-3317 Application 10/703,287 2 STATEMENT OF THE CASE Claims 1-11, 13-28, and 30-43 are pending (App. Br. 2). Claims 1-11, 13-28, 30, and 32-43 are on appeal (id.).2 Claims 1 and 13 are representative and read as follows: 1. A patient-adjustable urinary incontinence device, comprising: a pillow, configured for implantation adjacent a patient’s urethra; a reservoir, configured for implantation within said patient, wherein the reservoir is in fluid communication with the pillow; and a control mechanism which allows the patient to add and remove fluid from said pillow and said reservoir, to selectively enlarge said pillow to apply pressure against said urethra to inhibit urination. 13. A method of controlling urinary incontinence, comprising: providing a device comprising a pillow and a control element just under the skin, wherein said pillow is located adjacent the urethra; adjusting the pressure applied by said pillow to said urethra by activating said control element through the skin to selectively inhibit urination. The Examiner applies the following documents in rejecting the claims: Porter US 6,102,848 Aug. 15, 2000 Simon et al. US 5,090,424 Feb. 25, 1992 2 At Oral Argument Appellant stated that claim 31 was withdrawn from appeal. When prosecution resumes in this case claim 31 should therefore be canceled. See Ex Parte Ghuman, http://www.uspto.gov/web/offices/dcom/bpai/prec/rm081175.pdf, slip op. at 5-6 (Bd. Pat. App. & Int. May 1, 2008) (precedential). Appeal 2008-3317 Application 10/703,287 3 The following rejections are before us for review: Claims 1-3, 6-11, 13-17, 19-24, 26-28, 30, and 32-43 stand rejected under 35 U.S.C. § 102(b) as being anticipated by Porter (Ans. 3-5). Claims 1 and 5 stand rejected under 35 U.S.C. § 102(b) as being anticipated by Simon (Ans. 5-6). Claims 4, 18, and 25 stand rejected under 35 U.S.C. § 103(a) as being obvious in view of Porter (Ans. 6). ANTICIPATION -- PORTER ISSUE The Examiner contends that Porter anticipates claim 1 because it discloses a device having “a pillow (12) capable of being implanted in a urethra (102), [a] reservoir (14) in fluid communication with the pillow (12) capable of being implanted in a patient[,] and [a] control mechanism (16) that allows a patient to add and remove fluid from the pillow and reservoir to inhibit urination” (Ans. 3-4). The Examiner further contends that Porter discloses all of the features in claims 2-3, 6-11, 13-17, 19-24, 26-28 and 30 and 32-43. Appellant contends that the Examiner erred in finding that Porter anticipates claim 1 because Porter’s device is not implantable (App. Br. 4-6). Appellant submits additional separate arguments with respect to the remaining claims (see App. Br. 6-14). The issue with respect to this rejection, then, is whether the Examiner erred in finding that Porter meets all of the limitations recited in the rejected claims. Appeal 2008-3317 Application 10/703,287 4 FINDINGS OF FACT 1. Claim 1 recites a patient-adjustable urinary incontinence device. The device has a pillow and a reservoir which are in fluid communication with each other. Claim 1 requires the pillow to be “configured for implantation adjacent a patient’s urethra,” and the reservoir to be “configured for implantation within said patient.” The device also has a control mechanism that allows the patient to add and remove fluid from the pillow and reservoir, to selectively enlarge the pillow and thereby apply pressure against the urethra to inhibit urination. Independent claim 20 recites a similar device: 20. A patient-adjustable urinary incontinence device comprising: a pillow, configured for implantation adjacent a patient's urethra; a patient-adjustable valve for draining the pillow; and a patient-adjustable pump for filling the pillow to inhibit urination. 2. Porter discloses a “urethral valve system [which] includes a valve balloon that inflates to provide continence and in certain embodiments deflates to allow urine to flow around the balloon” (Porter, col. 1, ll. 44-47). Porter states that “[a] small-diameter catheter is attached to the balloon and extends through the urethra to provide access to the balloon from outside the body. When the urine flows around the balloon, the urine also flows around the catheter and through the urethra” (id. at col. 1, ll. 47-51). 3. Figures 1A and 1B of Porter, reproduced below, are, respectively, “an illustration of a first embodiment of a urethral valve system constructed in accordance with the invention providing continence,” and “an illustration of Appeal 2008-3317 Application 10/703,287 5 the first embodiment of the urethral valve system allowing urination” (Porter, col. 2, ll. 29-34): The figures show: [A] urethral valve system 10 [which] includes a valve balloon 12 that is positioned in a bladder 100 to obstruct the flow of urine from the bladder. The diameter of the balloon 12 must be large enough essentially to seal the neck 101 of the bladder 100 and prevent urine flow from the bladder. A small-diameter catheter 14 attaches to the balloon 12 and extends through the urethra 102 and past the meatus 104. A check valve 16, which is closed to retain within the system the substance, such as air or water, that is used to inflate the balloon 12, extends slightly below the meatus 104. A pump 18, which is removably attached to the check valve 16, provides, for example, the water to the system. When the balloon is inflated, as depicted in FIG. 1A, the balloon provides continence. (Porter, col. 3, ll. 10-24.) Porter discloses that, “[f]or urination, a user deflates the balloon 12 through the check valve 16. As depicted by the arrows in FIG. 1B, when the balloon 12 is deflated urine flows around the balloon, through the neck 101 Appeal 2008-3317 Application 10/703,287 6 of the bladder 100, through the urethra 102 and around catheter 14, and out of the meatus 104” (id. at col. 3, ll. 25-29). 4. Porter discloses that “[a]s depicted FIG. 1B, the pump 18 captures the water and retains it for use in re-inflating the balloon 12. The user may instead release the water from the system 10, and use fresh water to re- inflate the balloon” (Porter, col. 3, ll. 36-39). 5. Porter discloses that the “balloon 12 and the catheter 14 may be made of any inflatable, non-reactive material, such as silicon. The catheter 14 has a small diameter since urine flows around the catheter rather than through the catheter” (Porter, col. 3, ll. 49-52). 6. In another embodiment, Porter discloses that a valve 954 controlling urine flow can be placed in the bladder in conjunction with a pillow blocking the neck of the bladder (Porter, col. 5, ll. 44-64, Figures 10A and 10B). PRINCIPLES OF LAW “It is well settled that “[t]o anticipate a claim, a prior art reference must disclose every limitation of the claimed invention, either explicitly or inherently.” In re Schreiber, 128 F.3d 1473, 1477 (Fed. Cir. 1997). It is also well settled that “[a] patent applicant is free to recite features of an apparatus either structurally or functionally.” Id. at 1478. However, “‘[f]unctional’ terminology may render a claim quite broad . . .[;] a claim employing such language covers any and all embodiments which perform the recited function.” In re Swinehart, 439 F.2d 210, 213 (CCPA 1971)(emphasis added); see also Schreiber, 128 F.3d at 1478-79 (holding that a prior art apparatus meeting all claimed structural limitations was anticipatory because it was inherently capable of performing the claimed function). Appeal 2008-3317 Application 10/703,287 7 Moreover, as stated in In re Best, 562 F.2d 1252, 1255 (CCPA 1977) (quoting In re Swinehart, 439 F.2d at 212-13): [W]here the Patent Office has reason to believe that a functional limitation asserted to be critical for establishing novelty in the claimed subject matter may, in fact, be an inherent characteristic of the prior art, it possesses the authority to require the applicant to prove that the subject matter shown to be in the prior art does not possess the characteristic relied on. During examination, the PTO must interpret terms in a claim using “the broadest reasonable meaning of the words in their ordinary usage as they would be understood by one of ordinary skill in the art, taking into account whatever enlightenment by way of definitions or otherwise that may be afforded by the written description contained in the applicant’s specification.” In re Morris, 127 F.3d 1048, 1054 (Fed. Cir. 1997). This broad claim interpretation is proper at the examination stage because “during patent prosecution when claims can be amended, ambiguities should be recognized, scope and breadth of language explored, and clarification imposed.” In re Zletz, 893 F.2d 319, 321 (Fed. Cir. 1989). ANALYSIS Appellant’s arguments do not persuade us that the Examiner erred in concluding that Porter’s device meets all of the limitations in claims 1 and 20. Porter’s device has a valve balloon 12 that is positioned in the bladder such that it seals the neck of the bladder, preventing urine flow from the bladder (FF 3). Appellant does not dispute that Porter’s balloon meets the requirement in claims 1 and 20 for a pillow, but instead argues that Porter’s Appeal 2008-3317 Application 10/703,287 8 device does not meet the limitation requiring the pillow to be “configured for implantation” (App. Br. 4). Specifically, Appellant argues that, rather than using the term “implant,” Porter describes its device as being inserted through the urethra and bladder (id. at 5). Appellant urges that “[a]n implantable device and an insertable device are distinct to a person with ordinary skill in the art,” and that an “insertable device is not inherently implantable. An implantable device is subject to much greater FDA regulation and must meet heightened standards with respect to issues related to biocompatibility. Therefore, Porter does not describe an implantable pillow and is not inherently implantable” (id.). The Examiner responds that “three general definitions of ‘implant’ are to ‘to fix or set securely or deeply’, ‘to set in living tissue’, or ‘something implanted in tissue.’ See Merriam Websters [sic] Collegiate Dictionary, 10th Edition, 1996” (Ans. 7). Moreover, the Examiner argues, Appellant has “not clearly set forth in the[] specification, a definition of ‘implant’ that contradicts the plain meaning of that term” (id.). The Examiner concludes Porter’s pillow is encompassed by the claims when given their broadest reasonable interpretation consistent with the Specification because “[t]he insertable channel-blocking device of Porter is inherently implantable according to any of the three definitions of ‘implant’ described above” (id.). Given the definitions advanced by the Examiner, and the absence of direct evidentiary support for Appellant’s assertion that inserted devices would necessarily be recognized to be distinct from implanted devices, we find that Appellant has not shown that the language of claims 1 and 20 excludes Porter’s devices. Appeal 2008-3317 Application 10/703,287 9 Rather, we agree with the Examiner that it was reasonable to conclude that Porter’s balloon 12 is encompassed by claims 1 and 20. Specifically, because Porter’s balloon is placed in the patient’s bladder adjacent to the urethra, and is therefore set in living tissue, Porter’s device meets the general definition of “implant” advanced by the Examiner. We therefore agree that Porter’s balloon 12 meets the recitation in claims 1 and 20 of “a pillow, configured for implantation adjacent a patient’s urethra.” Appellant argues that, given the Specification’s disclosure of deploying the device “through a surgical cut or incision,” one of ordinary skill in the art viewing claims 1 and 20 in light of the Specification would interpret the device as one “that is configured to be implanted into a patient during a surgical procedure” (Reply Br. 2). We are not persuaded by this argument. “[W]hile it is true that claims are to be interpreted in light of the specification and with a view to ascertaining the invention, it does not follow that limitations from the specification may be read into the claims.” Sjolund v. Musland, 847 F.2d 1573, 1581 (Fed. Cir. 1988). Thus, “[a]bsent claim language carrying a narrow meaning, the PTO should only limit the claim based on the specification or prosecution history when those sources expressly disclaim the broader definition.” In re Bigio, 381 F.3d 1320, 1325 (Fed Cir. 2004). In the instant case, Appellant points to no specific definition of “implant” in the Specification that supplants or modifies the general definitions advanced by the Examiner. Moreover, Appellant points to no structural feature of Porter’s device that would render it incapable of being implanted by a surgical procedure. We therefore find that Appellant has not met the burden of rebutting the Examiner’s reasonable conclusion that Appeal 2008-3317 Application 10/703,287 10 Porter’s device meets the limitation requiring the pillow to be “configured for implantation.” Appellant argues that the Examiner erred in concluding that the term “reservoir” in claim 1 encompasses Porter’s catheter 14 (see FF 3) because the catheter “is so thin that urine is able to flow around the catheter while the catheter is inserted in the urethra. . . . Such a thin catheter would not be understood to be a reservoir by a person with ordinary skill in the art” (App. Br. 5 (citing Porter, col. 3, ll. 50-52)). The Examiner contends that Porter’s catheter 14 is encompassed by claim 1 because “it is a fluid reservoir (receptacle or chamber for storing a fluid) that is capable of holding pneumatic fluid between valve (16) and balloon/pillow (12)” (Ans. 8). We agree with the Examiner that the term “reservoir’ in claim 1 encompasses Porter’s catheter 14. As required by claim 1, Porter’s catheter is in fluid communication with the balloon (FF 3). Moreover, because Porter’s catheter holds fluid, and because that fluid is held for use within the balloon, we agree that the catheter can be considered to be a reservoir in relation to the balloon. Appellant argues that one of ordinary skill in the art reading claim 1 in light of the Specification would read “reservoir” to mean “a receptacle or chamber for storing a fluid that holds a sufficient amount of fluid to allow for inflation and deflation of the pillow” (Reply Br. 2). We are not persuaded by this argument. We note that, given its relative size, Porter’s catheter likely does not hold a large amount of fluid. However, claim 1 does not require the reservoir to hold any particular amount of fluid, much less an amount Appeal 2008-3317 Application 10/703,287 11 sufficient to allow the pillow to be inflated or deflated. As discussed above, it is improper to import limitations from the Specification into the claims. See Sjolund, 847 F.2d at 1581; Bigio, 381 F.3d at 1325. Thus, we agree with the Examiner that Porter’s device has both the pillow and reservoir recited in claim 1. Because check valve 16 is used as a control mechanism to selectively enlarge balloon 12 and apply pressure to the urethra to prevent urination (FF 3, 4), we also agree with the Examiner that Porter’s device has the claimed control mechanism. Therefore, because Porter’s device meets all of the limitations of claim 1, we affirm the Examiner’s anticipation rejection of that claim. Because they were not argued separately, claims 2, 3, 7-10, 32, and 33 fall with claim 1. 37 C.F.R. § 41.37(c)(1)(vii). Appellant reiterates their arguments regarding implantability with respect to claims 34 and 35, which depend from claim 1 (App. Br. 7). As discussed above, we do not find those arguments persuasive, and therefore affirm the Examiner’s rejection of those claims. With respect to claim 20, Porter’s device has the implantable pillow (balloon 12), patient-adjustable valve for draining the pillow (check valve 16), and patient-adjustable pump (pump 18) recited in that claim (see FF 3, 4). We therefore agree with the Examiner that Porter meets all of claim 20’s limitations, and affirm the Examiner’s anticipation rejection of claim 20. Because they were not argued separately, claims 21-23, 27, 40, and 41 fall with claim 20. 37 C.F.R. § 41.37(c)(1)(vii). Appellant reiterates their arguments regarding implantability with respect to claims 42 and 43, which depend from claim 20 (App. Br. 13-14). Appeal 2008-3317 Application 10/703,287 12 As discussed above, we do not find those arguments persuasive, and therefore affirm the Examiner’s rejection of those claims. Claim 6 recites “[t]he device of Claim 1, wherein said control mechanism comprises a bulb.” Appellant argues that the Examiner erred in finding that Porter’s pump 18 can be considered a bulb because Porter does not use the term “bulb” when describing that device, but instead only uses the term “pump” (App. Br. 6). Appellant argues that because a “pump” encompasses numerous different types of pump, a bulb being only one type of manually actuated pump, Porter does not anticipate claim 6 (id.). Appellant reiterates this argument with respect to claim 24 (id. at 11-12). We are not persuaded by these arguments. Because Porter discloses its pump 18 as having an oval shape that increases in size when it receives fluid from the balloon 12 (FF 3), we agree with the Examiner that it was reasonable to conclude that Porter’s pump is encompassed by the term “bulb.” We therefore affirm the Examiner’s anticipation rejection of claims 6 and 24. Claim 11 recites “[t]he device of Claim 1, wherein said control mechanism is configured for implantation just under the skin.” Appellant argues that Porter does not anticipate claim 11 because its valve is not inherently implantable just under the skin (App. Br. 7). Rather, Appellant argues, “Porter’s valve is designed to remain outside the body and is not designed to be the correct size and shape for implantation or to be biocompatible. Therefore, Porter's valve is not inherently implantable and does not anticipate claim 11” (id.). Appellant reiterates this argument with respect to claim 28 (id. at 12-13). Appeal 2008-3317 Application 10/703,287 13 The Examiner responds that “there is no language in the claims that indicate what structure configured for implantation just under the skin entails. It is the interpretation of the examiner that Porter’s control mechanism is capable of being implanted even if it is not optimized for that purpose and could cause complications if used in that manner” (Ans. 9). We agree with the Examiner that Porter’s device meets the limitations of claims 11 and 28. Those claims are apparatus claims, not method claims. Those claims therefore do not require the device to be actually implanted as claimed; they only require the device to be capable of such implantation. While Porter’s valve 16 is used outside the body (FF 3), Appellant points to no specific disclosure in Porter to support their assertion that the valve is incapable being implanted in accordance with the claim, particularly given the relatively broad definition of “implant” discussed above. Given the definition of “implant” and Porter’s description of its valve, a preponderance of the evidence supports the Examiner’s finding of anticipation. We therefore affirm the Examiner’s rejection of claims 11 and 28. Claim 13 reads as follows: 13. A method of controlling urinary incontinence, comprising: providing a device comprising a pillow and a control element just under the skin, wherein said pillow is located adjacent the urethra; adjusting the pressure applied by said pillow to said urethra by activating said control element through the skin to selectively inhibit urination. Appellant argues that the Examiner erred in finding that Porter anticipates claim 13 because Porter does not provide a pillow and control Appeal 2008-3317 Application 10/703,287 14 element just under the skin (App. Br. 8), and because sitting on the valve and pump would not “place the valve and pump just under the skin such that the valve and pump can be activated through the skin as required by claim 13. Porter’s valve and pump, even if sat upon, would still be activated outside the body, and not through the skin as required by claim 13” (id. at 9). The Examiner responds that “Porter teaches a control element (valve)(954) that is located within a bladder (100) which is under the skin (fig. 10A)(col. 5, line 8-col. 6, line 29). Applicants have not clearly set forth in their specification, a definition of ‘just under the skin’ that contradicts the plain meaning of that term” (Ans. 9). We agree with Appellant that Porter does not meet all of the limitations of claim 13. Even if one were to consider Porter’s bladder- implanted valve 954 to be implanted just under the skin, that valve controls urine flow, and does not adjust the pressure within the bladder-blocking balloon as required by claim 13 (FF 6). Also, in contrast to claim 13’s requirement of “activating said control element through the skin,” Porter’s control elements, valve 16 and pump 18, act through the urethra (FF 3, 4). Because we agree with Appellant that Porter does not meet all of claim 13’s limitations, we reverse the Examiner’s anticipation rejection of claim 13 and its dependent claims 14-19. Claims 30 and 36-39 also depend, directly or ultimately, from claim 13. However, each of those claims recites a “device” rather than a method as recited in claim 13. Nonetheless, because of their dependency on claim 13, we interpret claims 30 and 36-39 as reciting methods. Moreover, because they depend from claim 13, claims 30 and 36-39 also require all of the steps of claim 13’s Appeal 2008-3317 Application 10/703,287 15 method. We therefore reverse the Examiner’s anticipation rejection of those claims, for the reasons discussed above with respect to claim 13. Claim 26 recites “[t]he device of Claim 20, wherein said valve comprises a one-way valve directed contrary to the flow direction.” Appellant argues: The examiner asserted that Porter anticipated claim 26 because “Porter teaches a check valve.” (Final Office Action, page 5). However, Porter does not teach a check valve directed contrary to the flow direction (Porter, Specification, column 3, lines 17-19, 21-23, 25-27 and 44-46). Ordinarily, check valves are installed in the same direction as the flow. Because Porter does not specify the orientation of the check valve in the apparatus, it must be assumed that the check valve is installed in the flow direction. Therefore, Porter does not teach a one- way valve directed contrary to the flow direction and does not anticipate claim 26. (App. Br. 12.) The Examiner responds: [T]he plain meaning of “check valve” is “a valve that permits flow in one direction only[.]” See Merriam Websters [sic] Collegiate Dictionary, 10th Edition, 1996. As interpreted herein, “flow direction” is considered to be from balloon/pillow (12) towards valve (16). The valve prevents fluid from escaping the system and is thus directed against the flow direction (see col. 3, lines 11-24). Applicants have not clearly set forth in their specification, a definition of “check valve” or “flow direction” that contradicts the plain meaning of these terms. (Ans. 9-10.) We agree with Appellant that Porter does not meet the limitations of claim 26. We note that Porter describes its valve 16 as a check valve (FF 3). However, Porter discloses that the water captured by the pump 18 upon deflation of balloon 12 can be reused to inflate the balloon (FF 4). Appeal 2008-3317 Application 10/703,287 16 For that action to occur, the water would have to flow back through valve 16 in the opposite direction of the flow that occurred when the balloon deflated. Because valve 16 must therefore accommodate two directions of fluid flow, we do not agree with the Examiner that Porter’s valve is “a one- way valve directed contrary to the flow direction” as required by claim 26. We therefore reverse the Examiner’s rejection of claim 26 as anticipated by Porter. In summary, we affirm the Examiner’s rejection of claims 1-3, 6-11, 20-24, 27, 28, 32-35, and 40-43 under 35 U.S.C. § 102(b) as being anticipated by Porter. However, we reverse the Examiner’s anticipation rejection of claims 13-17, 19, 26, 30, and 36-39. ANTICIPATION -- SIMON ISSUE Claims 1 and 5 stand rejected under 35 U.S.C. § 102(b) as being anticipated by Simon (Ans. 5-6). The Examiner cites Simon as disclosing “a pillow (23') for implanting in a urethra, [a] reservoir (10) in fluid communication with the pillow (23') for implantation in a patient; and [a] control mechanism (24, 25) that allows a patient to add and remove fluid from the pillow and reservoir to inhibit urination” (Ans. 5). The Examiner finds that Simon meets the limitations of claim 5 because “the control mechanism (24, 25) is structurally and functionally a duckbill valve” (id. at 6). Appellant argues that Simon does not anticipate claims 1 and 5 because its device is a urethral plug that is inserted into the user’s urethra Appeal 2008-3317 Application 10/703,287 17 rather than being implanted (App. Br. 14). Appellant further argues that Simon’s device lacks the reservoir recited in claim 1 (id. at 15). FINDINGS OF FACT 7. Claim 5 recites “[t]he device of Claim 3, wherein said valves comprise duckbill valves.” Claim 3 in turn recites “[t]he device of Claim 1, wherein said control mechanism comprises check valves.” 8. Simon discloses “urethral plug which can be inserted into the urethra to restrict the flow of urine through the urethra and is used by a person suffering from urinary incontinence” (Simon, col. 1, ll. 6-10). 9. Figure 1 of Simon, reproduced below, “is a cross-sectional view showing the urethral plug in a deflated insertable configuration” (Simon, col. 3, ll. 54-55): The figure shows [U]rethral plug 10 . . . [with] bellows 11 [which] defines cavity 21, and is used to transport fluid 32 contained in cavity 21 through a check valve 22 which is located within meatal plate 13. The bellows is made from a material which makes it conformable to the body and comfortable for the patient when Appeal 2008-3317 Application 10/703,287 18 the urethral plug is in place. The fluid 32 is transported to cavity 23 located within plug 10 becoming fluid 33. (Simon, col. 5, ll. 9-16.) 10. Figure 2 of Simon, reproduced below, “is a cross-sectional view showing the urethral plug in an inflated configuration” (Simon, col. 3, ll. 56- 57): The figure shows cavity 23' in an inflated state that prevents passage of urine through the urethra. ANALYSIS We agree with the Examiner that Simon’s device meets all of the limitations of claims 1 and 5. As required by claim 5, the inflated cavity 23' can be considered to be the pillow recited in claim 1. Given the definition of “implant” discussed above, and the device’s disclosed capacity to be implanted within the urethra (FF 8, 9), we also agree with the Examiner that the cavity/pillow 23' meets the limitation requiring it to be configured for implantation. Appeal 2008-3317 Application 10/703,287 19 Moreover, while we note that reference number 10 refers to the entire device (FF 9), we also note that the channel between cavity 23' and cavity 21 is shown as containing fluid (FF 9, 10). We therefore agree with the Examiner that the channel can be considered to be a reservoir of fluid for the device. We also agree with the Examiner that bellows 11 meets claim 1’s limitation of a control mechanism that allows the patient to add and remove fluid from the pillow, and that the valve adjacent the bellows meets the requirement in claim 5 for a duckbill valve. For the reasons discussed above with respect to Porter, we do not find Appellant’s arguments regarding the implantation and reservoir limitations persuasive. We therefore affirm the Examiner’s rejection of claims 1 and 5 as anticipated by Simon. OBVIOUSNESS -- PORTER Claims 4, 18 and 25 stand rejected under 35 U.S.C. § 103(a) as being obvious in view of Porter (Ans. 6). Claims 4, 18, and 25 depend, respectively, from claims 1, 13, and 20, and recite that the control mechanism (claim 4), control element (claim 18), and pump (claim 25), of the parent claim “comprises a series of valves.” The Examiner found that “Porter does not teach a series of valves” but concluded that the claims would have been obvious under § 103 because “[o]ne of ordinary skill in the art would have recognized that redundant valves provide additional protection from backflow” (Ans. 6). Appellant presented no argument regarding this ground of rejection. Therefore, because we detect no deficiency in the Examiner’s prima facie Appeal 2008-3317 Application 10/703,287 20 case, we affirm the Examiner’s rejection of claims 4 and 25 as obvious over Porter. However, as discussed above, we agree with Appellant that Porter does not teach the limitation in claim 13 requiring the pressure in the pillow to be adjusted by activating the control element through the skin. Claim 18 depends from claim 13, and therefore also has that limitation. Because the Examiner has not explained why that limitation would have been obvious to a person of ordinary skill in the art, the Examiner has not established a prima facie case of obviousness with respect to claim 18. We therefore reverse the Examiner’s obviousness rejection of claim 18 over Porter. SUMMARY We affirm the Examiner’s rejection of claims 1-3, 6-11, 20-24, 27, 28, 32-35, and 40-43 under 35 U.S.C. § 102(b) as being anticipated by Porter. We reverse the Examiner’s anticipation rejection of claims 13-17, 19, 26, 30, and 36-39 over Porter. We affirm the Examiner’s rejection of claims 1 and 5 under 35 U.S.C. § 102(b) as being anticipated by Simon. We affirm the Examiner’s rejection of claims 4 and 25 under 35 U.S.C. § 103(a) as being obvious in view of Porter. We reverse the Examiner’s obviousness rejection of claim 18 over Porter. AFFIRMED-IN-PART Appeal 2008-3317 Application 10/703,287 21 cdc KNOBBE MARTENS OLSON & BEAR LLP 2040 MAIN STREET FOURTEENTH FLOOR IRVINE CA 92614 Copy with citationCopy as parenthetical citation