Ex Parte AXNESSDownload PDFPatent Trial and Appeal BoardAug 29, 201814780795 (P.T.A.B. Aug. 29, 2018) Copy Citation UNITED STA TES p A TENT AND TRADEMARK OFFICE APPLICATION NO. FILING DATE FIRST NAMED INVENTOR 14/780,795 09/28/2015 DAVID ROY AXNESS 24737 7590 08/31/2018 PHILIPS INTELLECTUAL PROPERTY & STANDARDS 465 Columbus A venue Suite 340 Valhalla, NY 10595 UNITED STATES DEPARTMENT OF COMMERCE United States Patent and Trademark Office Address: COMMISSIONER FOR PATENTS P.O. Box 1450 Alexandria, Virginia 22313-1450 www .uspto.gov ATTORNEY DOCKET NO. CONFIRMATION NO. 2013P00454WOUS 5955 EXAMINER MALAMUD, DEBORAH LESLIE ART UNIT PAPER NUMBER 3762 NOTIFICATION DATE DELIVERY MODE 08/31/2018 ELECTRONIC Please find below and/or attached an Office communication concerning this application or proceeding. The time period for reply, if any, is set in the attached communication. Notice of the Office communication was sent electronically on above-indicated "Notification Date" to the following e-mail address(es): patti. demichele@Philips.com marianne.fox@philips.com katelyn.mulroy@philips.com PTOL-90A (Rev. 04/07) UNITED STATES PATENT AND TRADEMARK OFFICE BEFORE THE PATENT TRIAL AND APPEAL BOARD Ex parte DAVID ROY AXNESS Appeal2017-008173 Application 14/780,795 1 Technology Center 3700 Before RICHARD M. LEBOVITZ, MICHAEL J. FITZPATRICK, and JOHN E. SCHNEIDER, Administrative Patent Judges. LEBOVITZ, Administrative Patent Judge. DECISION ON APPEAL This appeal involves claims directed to a portable defibrillator apparatus. The Examiner rejected the claims under 35 U.S.C. § 102 as anticipated and under 35 U.S.C. § 103 as obvious. We have jurisdiction under 35 U.S.C. § 6(b). The§ 102 and§ 103 rejections are reversed, but a new ground of rejection pursuant 37 C.F.R. § 41.50(b) is set forth under 35 U.S.C. § 112. STATEMENT OF CASE Claims 1-18 stand finally rejected by the Examiner as follows: 1 The Appeal Brief ("App. Br.") (dated Jan. 26, 2017) lists Koninklijke Philips N.V., Eindhoven, The Netherlands, as the real party in interest. Appeal2017-008173 Application 14/780,795 Claims 1-8 and 10-18 under 35 U.S.C. § I02(a)(l) as anticipated by Kaib (U.S. Pat. No. 6,169,387 Bl, issued Jan. 2, 2001) (hereinafter, "Kaib"). Examiner's Answer ("Ans.") 2. Claim 9 under 35 U.S.C. § 103 as obvious in view of Kaib. Ans. 4. There are three independent claims on appeal, claims 1, 8, and 10. Claim 1, which is representative of the rejected claims, is reproduced below: 1. A portable defibrillator apparatus comprising: - an electrode connector; - an operator actuated button; - an automatic power on circuit operable to actuate the defibrillator independent of the operator actuated button in response to an activation input; and - a controller in electrical communication with the electrode connector, the operator actuated button and the automatic power on circuit, the controller operable to detect a pattern of events and further operable to disable the automatic power on circuit in response to the detected pattern of events. CLAIM INTERPRETATION Before a claim can be compared to the prior art, it must be properly interpreted. For this reason, we begin with claim interpretation. The following disputed claim limitation appears in independent claims 1 and 8: a controller in electrical communication with the electrode connector, the operator actuated button and the automatic power on circuit, the controller operable to detect a pattern of events and further operable to disable the automatic power on circuit in response to the detected pattern of events Claim 1 (Appendix A). The meaning of the phrase "pattern of events" is at issue in the rejections. 2 Appeal2017-008173 Application 14/780,795 During patent examination proceedings, claim terms are given "the broadest reasonable meaning ... in their ordinary usage as they would be understood by one of ordinary skill in the art, taking into account whatever enlightenment by way of definitions or otherwise that may be afforded by the written description contained in the applicant's specification." In re Morris, 127 F.3d 1048, 1054 (Fed. Cir. 1997). Thus, we first look to the Specification for guidance as to the claim's meaning. The Specification does not provide a definition of "pattern of events," but it contains the following pertinent descriptions: In accordance with the objectives of the present invention, an improved automatic power on feature for a defibrillator is described which detects likely circumstances of inadvertent activation. The invention generally encompasses the detecting that there is a pattern of the defibrillator powering on automatically and not being used for therapy or diagnostic evaluation, e.g. a typical intended use. When the defibrillator detects such a pattern of events, the automatic power on feature is disabled. Then, when the AED [ automated external defibrillator] is used in a manner that indicates the AED has been handled by a human such as replacing the battery or powering on the AED manually via the on/off switch, the automatic power on feature is re-enabled. Spec 2:22-30 ( emphasis added). In operation, the automatic power on circuit senses a case opening, and sends an activation signal to controller 206. Controller 206 in turn retrieves data of previous activation events from memory 230. If the previous events combined with the present activation indicate that the activation is the latest in a pattern of events that indicates a repeated inadvertent activation, then controller 206 disables the automatic power on circuit and returns the defibrillator to a standby condition. Return to standby may occur after the timeout period has passed. 3 Appeal2017-008173 Application 14/780,795 In a preferred embodiment, the pattern of events is comprised of a series of detected automatic power on activations followed only by a timeout period deactivation, with no intervening activity. Spec. 5:7-16 (emphasis added). The above-reproduced disclosures in the Specification are consistent with a "pattern of events" requiring more than one event ("retrieves data of previous activation events," "series" of activations) to disable the automatic power on circuit, namely, the controller takes into account the past history of events to determine whether to disable the power of the defibrillator. The events must occur in a "pattern," which absent a definition in the Specification, we construe to have its ordinary and customary meaning of a "regular and intelligible form or sequence discernible in the way in whkh something happens or is done."2 The only detected event described in the Specification (see above) is of an "activation event" in which the defibrillator powers on. The Specification teaches that when more than one activation takes place, and the events are determined to be inadvertent, the "automatic power on circuit" is disabled to prevent further activation events from taking place and to conserve battery power (Spec. 2:10-12). Independent claims 1 and 8 are not expressly limited to an event being an inadvertent "activation" or "power on" event and we do not read preferred embodiments into a claim. "[A] particular embodiment appearing in the written description may not be read into a claim when the claim language is broader than the embodiment." SuperGuide Corp. v. DirecTV Enters., Inc., 358 F.3d 870, 875 (Fed. Cir. 2 https://en.oxforddictionaries.com/definition/pattern ( definition 1.2) ( accessed Aug. 7, 2018). 4 Appeal2017-008173 Application 14/780,795 2004). Thus, an "event" can be any action that the recited "controller" is programmed to detect. See Ans. 5. Independent claim 10 has a step of "automatically activating the defibrillator with the automatic power on circuit" and then a subsequent step of detecting a pattern of events "based on the automatically activating step." We interpret "based on" to mean that the detected events include the event of automatically activating the defibrillator. Thus, claim 10 is narrower in scope than claims 1 and 8, by requiring an event which is automatic activation. REJECTIONS BASED ON KAIB The Examiner found that Kaib describes a portable defibrillator apparatus comprising all the limitations of the independent claims. Ans. 2- 3. With respect to the limitation of the claims requiring "the controller operable to detect a pattern of events and further operable to disable the automatic power on circuit in response to the detected pattern of events," the Examiner cites Kaib, at column 5, lines 40-49, column 6, line 58----column 7, line 19, and col. 6, lines 40-49 and lines 58---61. Ans. 3, 5---6. Kaib teaches: The monitor-defibrillator 12 also makes adjustments for depletion of battery 18 capacity during periods when the device is not being used. When not in use ( such as when stored on a shelf or taken by the patient on a day's outing as a spare device) the monitor-defibrillator 12 will automatically power itself up at specified intervals and make adjustments to the battery run time to compensate for energy losses due to self-discharge of the battery and current draw of monitor-defibrillator 12 components when powered down. Kaib, col. 5, 11. 40-49. 5 Appeal2017-008173 Application 14/780,795 The above-reproduced passage refers to automatically powering up the defibrillator to adjust for battery run time. However, the disclosure does not expressly teach that the circuit for powering up is subsequently disabled as required by the rejected claims. Nonetheless, it would be understood from this disclosure that once the battery run time is adjusted, the monitor- defibrillator would be turned off. Thus, the detected "event," that led to disabling the automatic power on circuit would be the charged battery. However, Kaib only described the detection of one "event," not a "pattern of events" as required by the claims. The Examiner states that "it is clear [from the disclosure at column 5, lines 40-49] that the device of Kaib detects a pattern of self-discharge energy losses, and makes adjustments to the battery run time as a result." Ans. 6. However, the Examiner's statement does not adequately explain how the cited disclosure of Kaib describes or suggests detecting a "pattern of events" and turning off or disabling the circuit "in response to the detected pattern of events" as required by all the independent claims. The Examiner also cited the following disclosure from Kaib as describing the disputed limitation of the claims: Use of the slow charge mode permits the converter to be operated and a therapy pulse delivered to the patient when the battery 18 capacity is low. If during operation of the converter-defibrillator 19 in the slow charge mode the battery 18 voltage falls below a level at which the monitor-defibrillator 12 can reliably operate the converter-defibrillator 19, then the monitor-defibrillator 12 will deactivate the converter and evaluate the energy capability stored in the converter. Kaib, col. 6, 11. 40-49. 6 Appeal2017-008173 Application 14/780,795 The Examiner stated that this disclosure shows "it is the energy level of the system that is detected, and used to either continue system power or disable operating power, which constitutes the controller operable to detect a pattern of events and further operable to disable the automatic power on circuit in response to the detected pattern of events." Ans. 6. We agree with the Examiner that the "energy level" of the system can serve as a detected "event" in accordance with the claims. Kaib teaches that when the energy or voltage of the battery falls below a level, a "converter" is deactivated. The Examiner appears to have determined that the "converter" of Kaib corresponds to the "automatic power on circuit" recited in rejected claims 1, 8, and 10. The Examiner states that a pattern of events is detected (Ans. 6), which we understand to be a pattern of detecting more than one event of the battery voltage falling below a certain level. Even if it is assumed that the battery does fall below a certain level more than once, the disclosure at column 6, lines 40-49 relied upon by the Examiner, and reproduced above, does not teach that more than one event of this kind occurs and, that in response to the more than one event, the "automatic power on circuit" is disabled as required by the claims. Rather, it appears that only one event takes place when the battery voltage falls. The Examiner did not adequately explain how this disclosure describes or suggests detecting a "pattern of events" and turning off or disabling the circuit "in response to the detected pattern of events" as required by the all the independent claims. The Examiner also cited Kaib for its teaching of: If the monitor-defibrillator 12 determines that the battery 18 capacity has fallen below a level at which the system performance data is in danger of being corrupted then the 7 Appeal2017-008173 Application 14/780,795 monitor-defibrillator 12 will remove operating power. The removal of operating power will reserve the remaining battery 18 capacity for maintenance of the data storage/processor 22. Kaib, col. 6, 11. 58---64. The cited passage describes deactivating the operating power of the defibrillator when the energy level falls below a certain level, but does not teach detecting a series of events in which the battery level falls and then subsequently disabling the power. Even if the monitor-defibrillator detects the battery level at multiple times, the Examiner did not establish that the monitor-defibrillator is "operable to disable the automatic power on circuit in response to the detected pattern of events," namely, to consider the past history of the battery when disabling the converter. The last passage cited by the Examiner as describing the disputed limitation is as follows: Since this low level of battery 18 capacity is inadequate for reliable operation of the monitor-defibrillator, the best possible use of the remaining battery 18 capacity is to preserve the operational history of the device stored in the data storage/processor 22. When this state has been reached, the monitor-defibrillator will refuse to power up until connected to the patient base station 30. If required, the patient base station 30 will provide additional energy to the monitor-defibrillator 12 to insure proper functioning during this power up sequence. The patient base station will then retrieve the operational history from the monitor-defibrillator 12 and recharge the battery 18. Kaib, col. 6, 1. 65---col. 7, 1. 10. Here, Kaib teaches when the battery level is at a certain low amount, the "the monitor-defibrillator will refuse to power up." Thus, the power on circuit is disabled, but consistent with the disclosures quoted above, there is 8 Appeal2017-008173 Application 14/780,795 no evidence of a controller that is "operable to disable the automatic power on circuit in response to the detected pattern of events." Because the Examiner did not provide adequate evidence that the defibrillator described in Kaib is "operable to disable the automatic power on circuit in response to the detected pattern of events" as required by the claims, the anticipation rejection of claims 1 and 8 is reversed. Claim 10 also requires the detection of a pattern of events and disabling the automatic power on circuit based on this detecting step. Consequently, the anticipation rejection of claim 10 is also reversed. The anticipation rejection of dependent claims 2-7 and 11-18 is also reversed for the same reasons. The § 103 rejection of claim 9 based on Kaib is also reversed because dependent claim 9 incorporates all the limitations of claim 8 upon which it depends and the Examiner did not explain how Kaib renders the limitations of claim 8 obvious to one of ordinary skill in the art. NEW GROUND OF REJECTION As explained in the Claim Interpretation section above, the only event that we identified in the Specification is "activation" of the defibrillator when it powers on. Consistently, the Appeal Brief in describing the claimed subject matter only refers to the embodiment of "having an automatic power-on feature that turns the defibrillator on when the case lid is opened." Appeal Br. 4. The Appeal Brief further explains: If the case opening sensor or magnet fails or malfunctions, the defibrillator may undergo spurious activations and deactivations. If the carry case becomes overstuffed with accessories, the case surface may become distorted such that the case opening sensor may activate spontaneously. Repeated activations and deactivations may quickly deplete the battery. 9 Appeal2017-008173 Application 14/780,795 Id. The Appeal Brief describes the inventors as having solved this problem "by providing a circuit and controller which disables the automatic power-on feature upon sensing a pattern events, e.g. of inadvertent activations/deactivations, as described above." Id. at 5. Although the activation/deactivation events are the only events we identified in the Specification and which are described by Appellant in the Appeal Brief, the claims are not so limited, but cover any pattern of any events. The Specification, however, does not disclose how a controller would be made operable to detect any pattern of any events and then disable the automatic power on circuit in response to these events. 35 U.S.C. § 112(a) requires that the specification of a patent describe "the manner and process of making and using [ the invention], in such full, clear, concise, and exact terms as to enable any person skilled in the art to which it pertains ... to make and use the same." A specification is not enabling if a person of ordinary skill in the art would be unable to practice the invention without "undue experimentation." In re Wands, 858 F.2d 731, 737 (Fed. Cir. 1988). The only embodiment described in the Specification is of detecting an activation event. There is no direction or guidance on how to make the controller operable to detect other types of events and no working examples of other events; in contrast, claims 1 and 8 cover any pattern of any events, while only one pattern of specific events is enabled. See Wands, 858 F.2d at 737. Accordingly, pursuant to 37 C.F.R. § 4I.50(b), claim 1, 7, 8, and 9 are rejected under 35 U.S.C. § 112(a) as not being enabled for the full scope of the claims. This is a new ground of rejection. 10 Appeal2017-008173 Application 14/780,795 Claims 2---6 and 10-18 are not subject to this rejection because the claims require the detected events to comprise activation of the defibrillator which is enabled by the Specification. TIME PERIOD FOR RESPONSE This decision contains a new ground of rejection pursuant to 37 C.F.R. § 4I.50(b). Section 4I.50(b) provides "[a] new ground of rejection pursuant to this paragraph shall not be considered final for judicial review." Section 4I.50(b) also provides: When the Board enters such a non-final decision, the appellant, within two months from the date of the decision, must exercise one of the following two options with respect to the new ground of rejection to avoid termination of the appeal as to the rejected claims: (1) Reopen prosecution. Submit an appropriate amendment of the claims so rejected or new Evidence relating to the claims so rejected, or both, and have the matter reconsidered by the examiner, in which event the prosecution will be remanded to the examiner. The new ground of rejection is binding upon the examiner unless an amendment or new Evidence not previously of Record is made which, in the opinion of the examiner, overcomes the new ground of rejection designated in the decision. Should the examiner reject the claims, appellant may again appeal to the Board pursuant to this subpart. (2) Request rehearing. Request that the proceeding be reheard under §41.52 by the Board upon the same Record. The request for rehearing must address any new ground of rejection and state with particularity the points believed to have been misapprehended or overlooked in entering the new ground of rejection and also state all other grounds upon which rehearing is sought. 11 Appeal2017-008173 Application 14/780,795 Further guidance on responding to a new ground of rejection can be found in the MPEP § 1214.01. No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.F.R. § 1.136(a)(l). See 37 C.F.R. §§ 41.50(±), 4I.52(b). REVERSED; 37 C.F.R. § 4I.50(b) 12 Copy with citationCopy as parenthetical citation