Section 64-80-10 - Approved Sterilization Equipment and Monitoring Methods10.1. The operator shall ensure that the instruments used in body piercing are properly cleaned and packaged in sterilizer bags with a color change indicator and then sterilized in a sterilizer that: 10.1.a. Is sold as sterilizing equipment for medical instruments;10.1.b. Has been approved by the Department;10.1.c. Is used, cleaned and maintained to manufacturer's specifications; and10.1.d. Is tested at least monthly (unless otherwise specified by the manufacturer) by using a commercial biological monitoring system, utilizing Bacillus stearothermophilus spores, to assure that all microorganisms, including spores, have been destroyed. The operator shall maintain biological indicator test results for a period of three (3) years and make them available for inspection at all times.W. Va. Code R. § 64-80-10