177 Neb. Admin. Code, ch. 1, § 010

Current through May 7, 2024
Section 177-1-010 - CALIBRATION VERIFICATION

Calibration verification of evidentiary breath testing devices will be conducted by a maintenance officer every 40 days. It will consist of two verifications using different target values from wet bath simulator solutions, dry gas standards, or internal references .

010.01 When calibration verification checks are performed with certified wet bath simulator solutions, all approved evidentiary breath testing instruments shall be able to produce results within +/- 5% of the target value of the wet bath simulator solution.
010.02 When calibration verification checks are performed by means of approved internal standard(s), all approved evidentiary breath testing instruments shall be able to produce results within +/- 5% of the known target values of the standard(s). This tolerance shall be verified by the normal prescribed program and operation of the testing devices.
010.03 When calibration verification checks are performed with dry gas standards, all approved evidentiary breath testing instruments shall be able to produce results within +/5% of the target value of the dry gas standard after applying applicable altitude or topographic elevation correction factor supplied by the manufacturer. Such correction factor may be applied by the operator, the dry gas standard supplier, or by the instrument if pre-programmed.
010.04 If the instrument calibration cannot be verified to be accurate within the above cited limits, the instrument will be taken out of service and repairs made as set forth in section 011.

177 Neb. Admin. Code, ch. 1, § 010

Amended effective 6/19/2016.