Ariz. Admin. Code § 13-10-104

Current through Register Vol. 30, No. 19, May 10, 2024
Section R13-10-104 - Testing Procedures
A. Law enforcement agencies or individuals acting independently of law enforcement agencies who conduct alcohol concentration determinations by means of devices shall utilize a quality assurance program that is conducted by Quality Assurance Specialists or Operators and generate records of periodic maintenance. This quality assurance program shall include:
1. Criteria for ensuring the accurate and proper operation of devices by the regular performance of Calibration Checks and Quality Assurance Procedures as referenced in subsections (A)(2) and (A)(3);
2. Calibration Checks of devices that are performed within 31 days of each other as Standard Calibration Check Procedures or during a test sequence bracketing a duplicate breath test as Concurrent Calibration Check Procedures and recorded according to the requirements of the appropriate Quality Assurance Procedures set forth in Exhibits G-2, G-3, G-6, H-2 and I-2 or as approved by the Director according to R13-10-103(I). These checks shall indicate that the device is capable of determining the value of a standard alcohol concentration solution with an accuracy limit of ± 0.01 grams per 210 liters of breath or ± 10 percent, whichever is greater;
3. Quality Assurance Procedure checks of devices that are performed within 90 days of each other as Standard Quality Assurance Procedures or during a test sequence bracketing a duplicate breath test as Concurrent Quality Assurance Procedures, and recorded according to the requirements of the appropriate Quality Assurance Procedures set forth in Exhibits G-4, G-5, G-6, H-3, H-4 and I-2 or as approved by the Director according to R13-10-103(I). These checks shall indicate that the device is capable of proper operation and is functioning as required by the Quality Assurance Procedures for the device;
4. Standard alcohol concentration solutions, either liquid or gas, that are National Institute of Standards and Technology (NIST) traceable; and
5. Records of Calibration Checks, Quality Assurance Procedures and maintenance or repairs for each device in use.
B. An Operator shall utilize the Standard Operational Procedure approved by the Department for the device being operated in performing tests for the determination of alcohol concentration, as contained in Exhibits G-1, G-6, H-1 and I-1 or as approved by the Director according to R13-10-103(I).
C. Duplicate breath tests shall be administered at intervals of not less than five minutes nor more than 10 minutes. The results of both tests shall be within 0.020 alcohol concentration of each other. If the second test is not within 0.020 alcohol concentration of the first test, additional tests shall be administered until the results of two consecutive tests are within 0.020 alcohol concentration.

Ariz. Admin. Code § R13-10-104

New Section made by final rulemaking at 12 A.A.R. 1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2). Amended by final rulemaking at 26 A.A.R. 723, effective 6/1/2020.