Utah Admin. Code 432-950-11

Current through Bulletin No. 2024-21, November 1, 2024
Section R432-950-11 - Technical Specifications for Mammography
(1) The facility shall have available a phantom for use in the facility's ongoing quality control program.
(2) The facility shall evaluate image quality at least monthly using a phantom that produces measurements satisfactory to the supervising physician.
(3) The facility's evaluation of clinical images shall include the following:
(a) Positioning;
(b) Compression;
(c) Exposure level;
(d) Resolution;
(e) Contrast;
(f) Noise;
(g) Exam Identification;
(h) Artifacts.

Utah Admin. Code R432-950-11